FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 2580024 · Received May 15, 2012

Report

Report Number
1028232-2012-01208
Event Type
Injury
Date Received
May 15, 2012
Date of Event
March 9, 2012
Report Date
May 3, 2012
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFO THAT THIS RIGHT VENTRICULAR LEAD EXHIBITED NO CAPTURE AND NO SENSING AT MAXIMUM OUTPUTS. AN X-RAY NOTED THE LEAD HAD PULLED BACK INTO THE ATRIUM. THERE WERE NO ADVERSE PT EFFECTS REPORTED. ALL AVAILABLE INFO INDICATES THE DEVICE REMAINS IN SERVICE. AS NO FURTHER INFO CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFO BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization