HILL-ROM AIR-SHIELDS PHOTOTHERAPY UNIT
Report
- Report Number
- 2510954-1999-00004
- Event Type
- Malfunction
- Date Received
- January 6, 2000
- Date of Event
- June 11, 1999
- Report Date
- November 3, 1999
- Manufacturer
- HILL-ROM AIR-SHIELDS
- Product Code
- LBI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
IN 1999 A HILL-ROM AIR-SHIELDS SADDLE PHOTOTHERAPY UNIT (MODEL PTM78-2) WAS BEING USED IN CONJUNCTION WITH A HILL-ROM AIR-SHIELDS OVERHEAD WARMER (MODEL PM78-1). THE WARMER WAS BEING USED ON A NEONATE WITH AN OXYGEN TENT SURROUNDING THE INFANT. THE ATTENDING NURSE STATED THAT THE PHOTOTHERAPY UNIT HAD BEEN TURNED ON AND WAS RUNNING FOR APPROX 20 MINUTES BEFORE THE INCIDENT OCCURRED. THE NURSE SAID SHE SAW A HOT FRAGMENT ARC FROM THE TOP OF THE PHOTOTHERAPY UNIT OVER THE OVERHEAD WARMER AND LAND ONTO A CLOTH DIAPER SITTING ON A SHELF AT THE REAR OF THE WARMER. THE FRAGMENT BURNT A HOLE INTO THE CLOTH DIAPER BUT DID NOT IGNITE AN OPEN FLAME ON THE DIAPER. THE ATTENDING NURSE REMOVED THE PHOTOTHERAPY UNIT FROM SERVICE AND CONTACTED THE CLINICAL ENGINEERING DEPARTMENT WITHIN THE HOSPITAL. NO ONE WAS INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HILL-ROM AIR-SHIELDS PHOTOTHERAPY UNIT | PHOTOTHERAPY ATTACHMENT | LBI | HILL-ROM AIR-SHIELDS | PTM78-2 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |