I-STAT ACT KAOLIN CARTRIDGE
Report
- Report Number
- 2245578-2012-00333
- Event Type
- Malfunction
- Date Received
- May 14, 2012
- Date of Event
- March 23, 2012
- Report Date
- April 16, 2012
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- JBP
- PMA / PMN Number
- K023582
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). LOT# S11287, MFR DATE: 10/2011, EXPIRATION DATE: 03/28/2012.
ON (B)(6) 2012, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGES THAT YIELDED MULTIPLE QUALITY CHECK CODE (QCC) 31 DURING A PROCEDURE. CUSTOMER WAS USING LOTS S11321A & S11287. CUSTOMER STATES THAT HE TRIED RECOLLECTING THE SAMPLE FROM A LINE WITH A SYRINGE INTO CARTRIDGE 8 TIMES AND GOT QCC 31 ON ALL 8 CARTRIDGES. CAUSING A DELAY IN TREATMENT OF A PT SINCE THE I-STAT IS THE ONLY METHOD FOR TESTING ACT AT THE CUSTOMER SITE. ART: (B)(4) IN THE I-STAT SYSTEM MANUAL DEFINES QCC 31 AS "UNABLE TO POSITION SAMPLE" ERROR. THE ANALYZER DID NOT DETECT MOVEMENT OF SAMPLE ACROSS THE SENSORS. THIS COULD BE DUE TO COLT IN THE SAMPLE (ESPECIALLY IN NEONATES), TO NOT CLOSING THE SNAP CLOSURE ON THE CARTRIDGE, OR TO AN ABERRANT CARTRIDGE. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. AN INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT ACT KAOLIN CARTRIDGE | ACT KAOLIN CARTRIDGE | JBP | ABBOTT POINT OF CARE INC. | NA | S11321A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |