FDA Adverse Event Other Summary report: N

BACT/ALERT BPA AND BPN CULTURE BOTTLES

MDR report key: 2576515 · Received December 9, 2008

Report

Report Number
3002769706-2008-00009
Event Type
Other
Date Received
December 9, 2008
Date of Event
October 14, 2008
Report Date
December 3, 2008
Manufacturer
BIOMERIEUX, INC.
Product Code
JTA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION OF THIS INCIDENT HAS BEEN INITIATED BY BIOMERIEUX INC. THE COMMUNITY BLOOD CENTER INITIATED A TRANSFUSION REACTION INVESTIGATION ON (B)(6) 2008. THE REPORT STATED THAT "THE IMPLICATED UNIT WAS COLLECTED ON (B)(6) 2008 AND STERILITY SAMPLING WAS PERFORMED ON (B)(6) 2008. THE BACT/ALERT BOTTLES REMAINED NEGATIVE AND WERE REMOVED FROM THE INCUBATOR ON (B)(6) 2008. THE AEROBIC AND ANAEROBIC BOTTLES ALONG WITH THE ASSOCIATED FRESH FROZEN PLASMA UNIT WERE SENT TO THE REFERENCE LAB FOR CULTURES; NO BACTERIAL GROWTH WAS DETECTED." "THE GRAM STAIN AND CULTURE FROM THE PARTIALLY TRANSFUSED PLATELET PRODUCT SHOWED GRAM POSITIVE COCCI AND THE CULTURE GREW (B)(6)." "INVESTIGATION REVEALS THE BACT/ALERT SYSTEM FOR NOTIFICATION WAS FUNCTIONING AND REVIEW BY THE FDA INDICATED THAT ALL STANDARD OPERATING PROCEDURE HAD BEEN FOLLOWED."

Description of Event or Problem · 1

A RECIPIENT AT THE (B)(6) IN (B)(6), RECEIVED A PARTIALLY TRANSFUSED PLATELETS PHERESIS UNIT AND EXPERIENCED SEVERE DYSPNEA AND AN ELEVATED TEMPERATURES OF 40 DEGREES C.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BACT/ALERT BPA AND BPN CULTURE BOTTLES MICROBIAL GROWTH MONITOR JTA BIOMERIEUX, INC. 1020387 1020417

Patients

Seq Age Sex Outcome Treatment
1 Other