PERF ALLIANCE OK CITY OK 1
Report
- Report Number
- 1718850-2012-00047
- Event Type
- Other
- Date Received
- May 11, 2012
- Date of Event
- April 10, 2012
- Report Date
- April 11, 2012
- Manufacturer
- SORIN GROUP USA, INC.
- Product Code
- DWE
- PMA / PMN Number
- PREAMENDMENT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
SORIN GROUP RECEIVED A REPORT THAT THE BLOOD DIDN'T APPEAR TO CHANGE COLOR AS EXPECTED SO THE PERFUSIONIST WAS CONCERNED THAT THE OXYGENATOR MAY NOT HAVE BEEN OXYGENATING PROPERLY. UPON INSPECTION OF THE CIRCUIT, THE PERFUSIONIST FOUND THAT THE OXYGENATOR INLET AND OUTLET LINES HAD BEEN CONNECTED IN REVERSE. THE INLET LINE WAS CONNECTED TO THE OUTLET OF THE OXYGENATOR AND VICE VERSA. IN RESPONSE TO THE ISSUE, THE USER REPLACED THE ENTIRE CIRCUIT AND THE CASE WAS COMPLETED WITHOUT FURTHER ISSUES. THERE WAS NO INJURY TO THE PT. IT WAS LATER REPORTED THAT THE PT HAS BEEN RELEASED FROM THE HOSPITAL AND IS DOING WELL. THE OXYGENATOR AND CIRCUIT WERE RETURNED TO SORIN GROUP FOR EVAL. VISUAL INSPECTION OF THE CIRCUIT CONFIRMED THAT IT HAD BEEN ASSEMBLED INCORRECTLY. PERFORMANCE TESTING WAS PERFORMED ON THE RETURNED OXYGENATOR AND ON A CONTROL UNIT FROM INVENTORY. BOTH THE RETURNED UNIT AND CONTROL WERE TESTED ONCE WITH THE INLET/OUTLET LINES IN THE "CORRECT" ORIENTATION AND ONCE WITH THE LINES REVERSED. THIS TESTING FOUND THAT THE PERFORMANCE OF THE RETURNED UNIT WAS COMPARABLE TO THAT OF THE CONTROL UNIT AND THAT HAVING THE LINES CONNECTED IN REVERSE DID NOT AFFECT GAS TRANSFER OF EITHER DEVICE. BASED ON THE LABORATORY TESTING, THE RETURNED PRIMOX WAS FOUND TO BE FUNCTIONALLY SOUND AND CAPABLE OF MEETING A PT'S METABOLIC DEMANDS IN BOTH ORIENTATIONS. ALTHOUGH TESTING FOUND THAT THE MISSASSEMBLY DID NOT NEGATIVELY AFFECT THE DEVICE'S PERFORMANCE, IN ORDER TO PREVENT A RE-OCCURRENCE, THE PERSONNEL RESPONSIBLE FOR THE ASSEMBLY AND INSPECTION OF THIS LOT HAVE BEEN RETRAINED. NO FURTHER ACTION IS DEEMED NECESSARY. THE INCORRECT ORIENTATION CAUSED THE PERFUSIONIST TO CHANGE OUT THE DEVICE; THIS HOWEVER, DID NOT CAUSE PT INJURY. CHANGING OUT THE CIRCUIT TOOK OVER 3 MINUTES. THIS MEDWATCH REPORT IS BEING FILED AS A RESULT OF THE LENGTH OF TIME REQUIRED TO CHANGE THE UNIT OUT.
SORIN GROUP RECEIVED A REPORT THAT THE BLOOD DIDN'T APPEAR TO CHANGE COLOR AS EXPECTED SO THE PERFUSIONIST WAS CONCERNED THAT THE OXYGENATOR MAY NOT HAVE BEEN OXYGENATING PROPERLY. UPON INSPECTION OF THE CIRCUIT, THE PERFUSIONIST FOUND THAT THE OXYGENATOR INLET AND OUTLET LINES HAD BEEN CONNECTED IN REVERSE. THE INLET LINE WAS CONNECTED TO THE OUTLET OF THE OXYGENATOR AND VICE VERSA. IN RESPONSE TO THE ISSUE, THE USER REPLACED THE ENTIRE CIRCUIT AND THE CASE WAS COMPLETED WITHOUT FURTHER ISSUES. THERE WAS NO INJURY TO THE PT. IT WAS LATER REPORTED THAT THE PT HAS BEEN RELEASED FROM THE HOSPITAL AND IS DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERF ALLIANCE OK CITY OK 1 | CUSTOM PERFUSION PACK | DWE | SORIN GROUP USA, INC. | NA | 1207200043 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |