UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2050012-2012-01100
- Event Type
- Malfunction
- Date Received
- May 16, 2012
- Date of Event
- April 18, 2012
- Report Date
- April 18, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CAUSE OF THE LEAKING, PER THE FSE, WAS EITHER PINCHED UCTA DRAIN TUBING AND/OR AN OCCLUSION IN THE ACTIVE WASH STATION DRAIN. (B)(4).
THE CUSTOMER REPORTED TO CUSTOMER TECHNICAL SUPPORT (CTS) A LEAK ON THE BOTTOM OF THE CLOSE TUBE ACCESSION (CTA) ON THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. THE CUSTOMER STATED THAT THE LEAKING FLUID WAS WASH BUFFER II AND BIOLOGICAL FLUIDS/WASTE FROM THE DRAIN LINES. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LAB COAT, EYE PROTECTION, AND GLOVES. THE CUSTOMER DID NOT COME IN CONTACT WITH THE FLUID AND DID NOT SEEK MEDICAL ATTENTION. NO EXPOSURE (SPRAYED OR SPLASHED) TO MUCOUS MEMBRANES OR OPEN WOUNDS WERE REPORTED. NO DEATH, INJURY OR CHANGES TO PATIENT TREATMENT OR RESULTS ARE ASSOCIATED WITH THIS EVENT. CTS GENERATED SERVICE REQUEST TO INVESTIGATE THE EVENT. ON (B)(6) 2012 A FIELD SERVICE ENGINEER (FSE) PERFORMED THE FOLLOWING SERVICES ON THE INSTRUMENT: THE FSE FOUND LEAKING FROM THE OVERFLOW BOTTLE AND THE ACTIVE WASH STATION; THE BOTTLE WAS REMOVED AND CLEANED. DRAIN PUMP TUBING WAS REPLACED. ACTIVE WASH STATION WAS REMOVED AND REPLACED. A CARRYOVER TEST WAS PERFORMED AND PER FSE WAS "OK." DRAIN LINE WAS CHECKED FOR KINKS AND OCCLUSION NONE FOUND. SERVICE ACTIVITY WAS VERIFIED AND MET PUBLISHED PERFORMANCE SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |