FDA Adverse Event Injury Summary report: N

KARL STORZ BOWEL GRASPER

MDR report key: 2575136 · Received May 9, 2012

Report

Report Number
2575136
Event Type
Injury
Date Received
May 9, 2012
Date of Event
April 25, 2012
Report Date
May 4, 2012
Manufacturer
KARL STORZ
Product Code
GCJ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NC, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

THE PT IS A (B)(6), FEMALE, DIAGNOSED WITH APPENDICITIS AND ADMITTED TO WFBH FOR A LAPAROSCOPIC APPENDECTOMY ON (B)(6) 2012. DURING THE PROCEDURE A LAPAROSCOPIC BOWEL GRASPER WAS INTRODUCED THROUGH THE INFERIOR PORT IN AN ATTEMPT TO TAKE DOWN SEVERAL ADHESIONS THAT WERE LOCATED AROUND THE AREA OF THE UMBILICUS. ONE-HALF OF THE BOWEL GRASPER BROKE LOOSE AND A PORTION WAS LOST IN THE ABDOMINAL CAVITY. THE SURGEON COMPLETED THE APPENDECTOMY THEN BEGAN TO SEARCH FOR THE BROKEN PIECE OF THE BOWEL GRASPER. AN ADDITIONAL PORT WAS PLACED IN THE LEFT UPPER QUADRANT TO AID IN THE SEARCH FOR THE MISSING PIECE OF INSTRUMENT. THERE WAS DIFFICULTY IN LOCATING THE BROKEN PIECE SO A DECISION WAS MADE TO PLACE A HAND ASSIST PORT. THE HAND ASSIST PORT WAS PLACED IN THE STANDARD FASHION AND THE PIECE OF BROKEN INSTRUMENT WAS FOUND DEEP ON THE PELVIS, THE PIECE WAS REMOVED WITHOUT DIFFICULTY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KARL STORZ BOWEL GRASPER LAPAROSCOPIC BOWEL GRASPER GCJ KARL STORZ 33300 NA

Patients

Seq Age Sex Outcome Treatment
1 44 YR Required Intervention HAND PIECE TO GRASPER IS 33121 CLICKLINE