FDA Adverse Event Malfunction Summary report: N

OT VERIO PRO METER

MDR report key: 2574849 · Received May 15, 2012

Report

Report Number
3008382007-2012-02059
Event Type
Malfunction
Date Received
May 15, 2012
Report Date
April 23, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED (B)(4) 2012 AND EVALUATED BY PRODUCT ANALYSIS (PA) RESPECTIVELY ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER PASSED TESTING WITH NO FAULTS FOUND. THE METER WAS FOUND TO WORK PROPERLY. THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS. THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIOPRO METER READ INACCURATELY ERRATIC. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "52, 221, 137, 53, AND 56 MG/DL" WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR PRECISION TESTING, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO PRO METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3213839

Patients

Seq Age Sex Outcome Treatment
1