FDA Adverse Event Malfunction Summary report: N

INTRAMATIC LUX3 CONTRA ANGLE 25LHA

MDR report key: 2572542 · Received May 8, 2012

Report

Report Number
3003637274-2012-00025
Event Type
Malfunction
Date Received
May 8, 2012
Date of Event
April 1, 2011
Report Date
April 10, 2012
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS DID SHOW THAT THE HANDPIECE HAD A MEDIUM RESIDUE. THIS SHOWS THAT THE REPROCESSING IS NOT CARRIED OUT LIKE REQUESTED. THE BEARINGS HAVE BEEN GRINDING, VIBRATING AND COMING APART, WHICH IS THE RESULT OF A NORMAL WEAR PROCESS. THE HANDPIECE HAS BEEN >5 YEARS IN USE SINCE THE LAST SVC CHECK! THE USER INSTRUCTION STATES FOLLOWING: DUE TO INCREASE TORQUE AND SPEED, POORLY MAINTAINED, DAMAGED, OR MISUSED HANDPIECES CAN POTENTIALLY GENERATE FRICTIONAL HEAT CAPABLE OF CAUSING SERIOUS BURN INJURIES TO THE PT. THE FOLLOWING GUIDELINES ARE ESSENTIAL FOR THE SAFE OPERATION OF ELECTRIC MICROMOTORS HANDPIECES: TEST RUN HANDPIECE ATTACHMENT, INSPECTING FOR HEAT, EXCESSIVE NOISE OR VIBRATION. IMMEDIATELY STOP USING ANY SUSPECT HANDPIECE. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT THE HAND PIECE HEATED UP AND BURNED THE PT ON THE CHEEK TO THE SIZE OF A NICKEL. THE DENTIST DID NOT GIVE OR PRESCRIBE ANYTHING FOR THE BURN. PT WAS TOLD TO DO WARM SALT WATER RINSES AND DID HEAL WELL. DENTAL OFFICE DOES NOT RECALL THE NAME OF THE PT, HENCE ONLY GENDER IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTRAMATIC LUX3 CONTRA ANGLE 25LHA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LHA

Patients

Seq Age Sex Outcome Treatment
1