FDA Adverse Event Malfunction Summary report: N

XTREMEBEAM

MDR report key: 2571289 · Received September 7, 2010

Report

Report Number
3005775121-2010-00001
Event Type
Malfunction
Date Received
September 7, 2010
Date of Event
June 23, 2010
Report Date
July 23, 2010
Manufacturer
BFW INC.
Product Code
FST
PMA / PMN Number
K093386
Product Problem
Yes
Report Source
Distributor report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING WORK ORDER NOW INCLUDES THE ADDITION OF LOCK TITE ON IRIS-ADJUSTING POST TO ALL HEADLIGHTS WITH 100% INSPECTION OF UNITS PRIOR TO SHIPMENT.

Description of Event or Problem · 1

(B)(6). INCIDENT: IRIS-ADJUSTING POST FELL FROM FIBER OPTIC HEADLIGHT OPTIC INTO PATIENT'S BODY CAVITY. POST WAS RETRIEVED FROM PATIENT. HEADLIGHT HAS JUST BEEN RECEIVED FROM FACILITY ACTION TAKEN BY FACILITY: HOSPITAL HAS FILED AN INTERNAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XTREMEBEAM FIBER OPTIC HEADLIGHT FST BFW INC. BFW 9855

Patients

Seq Age Sex Outcome Treatment
1 Other