FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2569325 · Received May 10, 2012

Report

Report Number
2017865-2012-03482
Event Type
Malfunction
Date Received
May 10, 2012
Date of Event
February 6, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED ANALYSIS FOUND INTERNAL ABRASION AT 8CM TO 9CM FROM THE DISTAL TIP. EXTERNAL ABRASION NOTED AT 36.3CM TO 38.5CM FROM THE HE LIX END. (B)(4). COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1581/65 NA

Patients

Seq Age Sex Outcome Treatment
1