FDA Adverse Event
Other
Summary report: N
2246640-1999-00003
MDR report key: 256917
·
Received December 15, 1999
Report
- Report Number
- 2246640-1999-00003
- Event Type
- Other
- Date Received
- December 15, 1999
- Date of Event
- November 9, 1999
- Report Date
- December 15, 1999
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
PT UNDERWENT A POSTERIOR INTERBODY FUSION PROCEDURE IN 1999, IN WHICH THE ALLEGED MEDICAL DEVICE WAS USED. DURING THE PROCEDURE, THE SURGEON TIRED TO DISTRACT THE L4-L5 DISC SPACE WITH THE ALLEGED DISTRACTOR ASSEMBLY (ATTACHMENT 00-2114-15 AND SHAFT 00-2114-14). THE SURGEON ENCOUNTERED DIFFICULTY DISTRACTING THE DISC SPACE, DUE TO ITS REPORTED TIGHTNESS. AT ONE POINT, THE DISTRACTOR ATTACHMENT SHEARED OFF OF THE DISTRACTOR SHAFT AND BECAME TEMPORARILY LODGED IN THE DISC SPACE. THE BROKEN PIECE WAS SUBSEQUENTLY REMOVED. THE PT WAS REPORTED STABLE WITH NO SEQUELAE ON 11/16/99.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention |