FDA Adverse Event
Malfunction
Summary report: N
2243801-1999-00041
MDR report key: 256872
·
Received December 22, 1999
Report
- Report Number
- 2243801-1999-00041
- Event Type
- Malfunction
- Date Received
- December 22, 1999
- Date of Event
- November 1, 1999
- Report Date
- November 29, 1999
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
RETURN REASON: THE CUSTOMER REPORTED THE CONNECTOR CAME OFF EASILY DURING THE PROCEDURE. ANALYSIS DETERMINED: INVESTIGATION FOUND THAT THE THERMISTOR CONNECTOR HOUSING BECAME DISCONNECTED FROM THE EXTENSION TUBING DUE TO INSUFFICIENT ADHESIVE. UNIT WAS USED IN VIVO. THIS WAS NOT DETERMINED UNTIL ANALYSIS WAS COMPLETED. NO FURTHER INFO IS AVAILABLE ON THE PT'S STATUS. CORRECTIVE ACTION TAKEN: THE CORRECTIVE ACTION IS THAT MFG AND QA PERSONNEL HAVE BEEN NOTIFIED. BOTH THE FREQUENCY AND SEVERITY OF THIS TYPE OF INCIDENT WILL BE CLOSELY MONITORED TO DETERMINE IF FURTHER CORRECTIVE ACTION IS NECESSARY.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |