FDA Adverse Event
Other
Summary report: N
BARIL0A BARIATRIC BED
MDR report key: 2560541
·
Received April 26, 2012
Report
- Report Number
- 2182305-2012-00014
- Event Type
- Other
- Date Received
- April 26, 2012
- Date of Event
- March 22, 2012
- Report Date
- April 23, 2012
- Manufacturer
- JOERNS HEALTHCARE
- Product Code
- FNL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED TO THE MFR, BY DEALER, (B)(4), PER DEALER, ALL WIRES ON THE BED WERE EXPOSED AND CAUGHT ON FIRE WITH PT ON THE BED. THE PT WAS UNINJURED. FURTHER INVESTIGATION CONFIRMED NO PT INJURY. PER FACILITY, "THE WIRE INSIDE THE WALL BROKE AND CAME LOOSE INSIDE THE PLUG. WHEN THE BED WAS PLUGGED IN, IT SPARKED AND CAUSES A FIRE. THE INSULATION ON THE SPIRAL CORD WAS LOOSE AND HAD EXPOSED WIRES WHICH MADE IT DOUBLY DANGEROUS". PER FACILITY, THE ISSUE WAS WITH THE PLUG, NOT THE BED. UPON RECEIPT OF THE BED, PRELIMINARY INVESTIGATION INDICATES USER ABUSE. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BARIL0A BARIATRIC BED | HOSPITAL BED | FNL | JOERNS HEALTHCARE | BARIL0A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |