FDA Adverse Event
Other
Summary report: N
ZIMMER DRILL STOP KIT
MDR report key: 2560368
·
Received May 21, 2010
Report
- Report Number
- 2023141-2010-00003
- Event Type
- Other
- Date Received
- May 21, 2010
- Date of Event
- May 11, 2010
- Report Date
- May 20, 2010
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZA
- PMA / PMN Number
- JTF #92
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PART NO: (B)(4), LOT NO: 61482528, TOTAL QUANTITY PRODUCED: (B)(4) (MFG DATE 3/20/2010); DSKIT, (B)(4) (MFG DATE 3/26/2010); (B)(4) (MFG DATE 4/16/2010); (B)(4) (MFG DATE 3/17/2010) - NO EXP DATE. INCOMPLETE STERILIZATION COULD LEAD TO CROSS CONTAMINATION OF MULTIPLE PTS. TESTING ON A PROTOTYPE PART WITH SIMILAR FEATURES CAME BACK SHOWING THAT STERILITY MAY BE COMPROMISED AT THE LID-KIT INTERFACE LOCATION, IDENTIFIED BY THE MFR.
Description of Event or Problem · 1
AS IDENTIFIED BY THE MFR: INSTRUCTIONS FOR USE CONTAIN PRE-VACUUM STERILIZATION INSTRUCTIONS THAT MAY LEAD TO COMPROMISED STERILITY AT THE LID-KIT INTERFACE LOCATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZIMMER DRILL STOP KIT | DRILL STOP KIT | DZA | ZIMMER DENTAL | DSKIT | 61482528 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |