FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2559925 · Received April 20, 2012

Report

Report Number
2245578-2012-00321
Event Type
Malfunction
Date Received
April 20, 2012
Date of Event
January 25, 2012
Report Date
April 13, 2012
Manufacturer
ABBOTT POINT OF CARE, INC.
Product Code
MMI
PMA / PMN Number
K031739
Removal / Correction Number
APOC-2012-002
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ON (B)(4) 2012, THE INCIDENT WAS IDENTIFIED AND LINKED TO AN INVESTIGATION THAT WAS COMPLETED ON (B)(4) 2012 AND A DEFICIENCY WAS IDENTIFIED. (B)(4). DETAILS WERE PROVIDED WITH THE ACTION THAT WAS CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.

Description of Event or Problem · 1

ON (B)(6) 2012, ABBOTT POINT OF CARE WAS CONTACTED BY AN ABBOTT DISTRIBUTOR IN THE VETERINARY MARKET WHO'S CUSTOMER COMPLAINT IS THAT I-STAT TROPONIN CARTRIDGES ARE YIELDING A HIGHER THAN NORMAL RATE OF QUALITY CHECK CODE (QCC) 69, QCC 133 AND QCC 145. THE CUSTOMER STATES THAT 3 OF 10 FAILED. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE, INC. NA U11246

Patients

Seq Age Sex Outcome Treatment
1 NA