FDA Adverse Event Malfunction Summary report: N

INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 2559299 · Received May 4, 2012

Report

Report Number
9611451-2012-00279
Event Type
Malfunction
Date Received
May 4, 2012
Date of Event
April 11, 2012
Report Date
April 11, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT RT265 INFANT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKE HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED BREATHING CIRCUIT. FURTHER INSPECTION REVEALED THAT THE LUER PORT CAP WAS MISSING FROM THE PRESSURE LINE PORT, CONFIRMING THE REPORTED FAULT. THE PRESSURE TEST RESULT WAS OUTSIDE OF THE REQUIRED SPECIFICATION. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 120125. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED PRIOR TO LEAVING THE PRODUCTION LINE AND THOSE THAT FAIL ARE REJECTED. HOWEVER, WE ARE UNABLE TO DETERMINE WHETHER THE SUBJECT LUER PORT CAP FELL OFF AT THE PRODUCTION LINE OR UPON OPENING OF THE RT265 KIT AT THE USER'S FACILITY. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR INFANT BREATHING CIRCUITS IN THE PAST 12 MONTHS TO THE END OF APRIL 2012.

Description of Event or Problem · 1

A HOSPTIAL IN THE (B)(6) REPORTED THAT THE "WHITE CAP OF THE TUBING" OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT WAS MISSING. THIS WAS NOTICED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT265 120125

Patients

Seq Age Sex Outcome Treatment
1