UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2012-01405
- Event Type
- Malfunction
- Date Received
- May 3, 2012
- Date of Event
- April 9, 2012
- Report Date
- April 9, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKJ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- 2050012-04/05/2012-006C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER STATED THE PUDDLE WAS 6 INCHES LONG HOWEVER, HE WAS UNABLE TO DETERMINE THE QUANTITY.
THE CAUSE OF THE LEAK WAS A PLUGGED NRBC MIX CHAMBER. (B)(4).
THE CUSTOMER REPORTED A LEAK OF APPROXIMATELY 4 ML OF BLOODY FLUID THAT WAS CONTAINED IN THE AIR MIX TEMPERATURE CONTROL (AMTC) MODULE OVERFLOW TRAY ON THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER STATED THE PUDDLE WAS 6 INCHES LONG, HOWEVER, HE WAS UNABLE TO DETERMINE THE QUANTITY. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, EYE PROTECTION, AND LAB COAT AT THE TIME OF THE INCIDENT. PATIENT SAMPLES OR RESULTS WERE NOT REPORTED TO HAVE BEEN AFFECTED BY THE LEAK. NO INJURIES OCCURRED AND MEDICAL ATTENTION WAS NOT SOUGHT. THE CUSTOMER DID NOT REPORT EXPOSURE (SPLASHED OR SPRAYED) TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE CUSTOMER DID NOT REVIEW THE MSDS, HOWEVER, THE FACILITY DOES HAVE AN EXPOSURE CONTROL / RISK MANAGEMENT PLAN IN PLACE. ON THE DAY OF THE EVENT, A FIELD SERVICE ENGINEER (FSE) FOUND DEBRIS LODGED IN THE NRBC MIX CHAMBER. THE DEBRIS WAS RUBBER FROM THE SPECIMEN TUBE STOPPER. THE DEBRIS PREVENTED THE CHAMBER FROM DRAINING, WHICH CAUSED IT TO OVERFLOW INTO OVERFLOW TRAY UNDERNEATH THE AMTC MODULE. THE FSE REMOVED THE DEBRIS FROM THE CHAMBER TO CORRECT THE ISSUE. BLOOD, 6C CELL CONTROL, DILUENT, AND DXH CELL LYSE MAY BE PRESENT IN THE NRBC MIX CHAMBER DURING NORMAL OPERATION, AND DXH CLEANER DURING SHUTDOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | SPINNER, SLIDE, AUTOMATED | GKJ | BECKMAN COULTER, INC. | DXH 800 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |