FDA Adverse Event Malfunction Summary report: N

UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM

MDR report key: 2559254 · Received May 3, 2012

Report

Report Number
1061932-2012-01405
Event Type
Malfunction
Date Received
May 3, 2012
Date of Event
April 9, 2012
Report Date
April 9, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKJ
PMA / PMN Number
K081930
Removal / Correction Number
2050012-04/05/2012-006C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER STATED THE PUDDLE WAS 6 INCHES LONG HOWEVER, HE WAS UNABLE TO DETERMINE THE QUANTITY.

Additional Manufacturer Narrative · 1

THE CAUSE OF THE LEAK WAS A PLUGGED NRBC MIX CHAMBER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK OF APPROXIMATELY 4 ML OF BLOODY FLUID THAT WAS CONTAINED IN THE AIR MIX TEMPERATURE CONTROL (AMTC) MODULE OVERFLOW TRAY ON THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER STATED THE PUDDLE WAS 6 INCHES LONG, HOWEVER, HE WAS UNABLE TO DETERMINE THE QUANTITY. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, EYE PROTECTION, AND LAB COAT AT THE TIME OF THE INCIDENT. PATIENT SAMPLES OR RESULTS WERE NOT REPORTED TO HAVE BEEN AFFECTED BY THE LEAK. NO INJURIES OCCURRED AND MEDICAL ATTENTION WAS NOT SOUGHT. THE CUSTOMER DID NOT REPORT EXPOSURE (SPLASHED OR SPRAYED) TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE CUSTOMER DID NOT REVIEW THE MSDS, HOWEVER, THE FACILITY DOES HAVE AN EXPOSURE CONTROL / RISK MANAGEMENT PLAN IN PLACE. ON THE DAY OF THE EVENT, A FIELD SERVICE ENGINEER (FSE) FOUND DEBRIS LODGED IN THE NRBC MIX CHAMBER. THE DEBRIS WAS RUBBER FROM THE SPECIMEN TUBE STOPPER. THE DEBRIS PREVENTED THE CHAMBER FROM DRAINING, WHICH CAUSED IT TO OVERFLOW INTO OVERFLOW TRAY UNDERNEATH THE AMTC MODULE. THE FSE REMOVED THE DEBRIS FROM THE CHAMBER TO CORRECT THE ISSUE. BLOOD, 6C CELL CONTROL, DILUENT, AND DXH CELL LYSE MAY BE PRESENT IN THE NRBC MIX CHAMBER DURING NORMAL OPERATION, AND DXH CLEANER DURING SHUTDOWN.