VITROS CHEMISTRY PRODUCTS PHBR SLIDES
Report
- Report Number
- 1319809-2012-00045
- Event Type
- Malfunction
- Date Received
- May 3, 2012
- Date of Event
- March 19, 2012
- Report Date
- May 3, 2012
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- DLZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE INVESTIGATION DETERMINED THAT IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS WERE OBTAINED ON TWO VITROS 5,1 FS CHEMISTRY SYSTEMS. SERVICE ACTIONS WERE PERFORMED ON BOTH ANALYZERS, HOWEVER, THERE WAS NO EVIDENCE TO SUGGEST AN ANALYZER MALFUNCTION CONTRIBUTED TO THE EVENT. FOLLOWING THE SERVICE ACTIONS, ACCEPTABLE VITROS PHBR PERFORMANCE WAS OBSERVED ON BOTH ANALYZERS. HOWEVER, THE CUSTOMER CHOSE TO PUT AN ALTERNATE VITROS PHBR REAGENT LOT INTO USE. THE INVESTIGATION WAS UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE. HOWEVER, VITROS PHBR LOT 2537-0058-8464 COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR. THE ROOT CAUSE OF THIS EVENT IS UNKNOWN.
A CUSTOMER OBTAINED IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS ON TWO VITROS 5,1 FS CHEMISTRY SYSTEMS. A VALUE OF 20.32 UG/ML WAS OBTAINED FROM THE VITROS TDM PV II FLUID VERSUS THE EXPECTED VALUE OF 26.00 UG/ML. VALUES OF 78.95, 75.92, 73.28, 70.79, 42.11, AND 41.60 UG/ML WERE OBTAINED FROM THE VITROS TDM PV III FLUID VERSUS THE EXPECTED VALUE OF 55.93 UG/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SEVEN OF SEVEN MDRS FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS CHEMISTRY PRODUCTS PHBR SLIDES | IN-VITRO DIAGNOSTIC | DLZ | ORTHO-CLINICAL DIAGNOSTICS | 2537-0058-8464 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |