FDA Adverse Event Malfunction Summary report: N

VITROS CHEMISTRY PRODUCTS PHBR SLIDES

MDR report key: 2559106 · Received May 3, 2012

Report

Report Number
1319809-2012-00043
Event Type
Malfunction
Date Received
May 3, 2012
Date of Event
March 19, 2012
Report Date
May 3, 2012
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
DLZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS WERE OBTAINED ON TWO VITROS 5,1 FS CHEMISTRY SYSTEMS. SERVICE ACTIONS WERE PERFORMED ON BOTH ANALYZERS, HOWEVER THERE WAS NO EVIDENCE TO SUGGEST AN ANALYZER MALFUNCTION CONTRIBUTED TO THE EVENT. FOLLOWING THE SERVICE ACTIONS, ACCEPTABLE VITROS PHBR PERFORMANCE WAS OBSERVED ON BOTH ANALYZERS. HOWEVER, THE CUSTOMER CHOSE TO PUT AN ALTERNATE VITROS PHBR REAGENT LOT INTO USE. THE INVESTIGATION WAS UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE. HOWEVER, VITROS PHBR LOT 2537-0058-8464 COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR. THE ROOT CAUSE OF THIS EVENT IS UNKNOWN.

Description of Event or Problem · 1

A CUSTOMER OBTAINED IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS ON TWO VITROS 5,1 FS CHEMISTRY SYSTEMS. A VALUE OF 20.32 G/ML WAS OBTAINED FROM THE VITROS TDM PV II FLUID VERSUS THE EXPECTED VALUE OF 26.00 G/ML. VALUES OF 78.95, 75.92, 73.28, 70.79, 42.11, AND 41.60 G/ML WERE OBTAINED FROM THE VITROS TDM PV III FLUID VERSUS THE EXPECTED VALUE OF 55.93 G/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FIVE OF SEVEN MDR'S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS CHEMISTRY PRODUCTS PHBR SLIDES IN-VITRO DIAGNOSTIC DLZ ORTHO-CLINICAL DIAGNOSTICS 2537-0058-8464

Patients

Seq Age Sex Outcome Treatment
1