FDA Adverse Event Malfunction Summary report: N

RESTORE ADVANCED

MDR report key: 2558518 · Received May 3, 2012

Report

Report Number
3004209178-2012-02946
Event Type
Malfunction
Date Received
May 3, 2012
Report Date
April 8, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ID 377675, LOT# V005336, IMPLANTED: (B)(6) 2008, PRODUCT TYPE LEAD; PRODUCT ID 37752, SERIAL# (B)(4), PRODUCT TYPE RECHARGER; PRODUCT ID 3708260, SERIAL# (B)(4), IMPLANTED: (B)(6) 2008, PRODUCT TYPE EXTENSION; PRODUCT ID 3487A-45, LOT# V078945, IMPLANTED: (B)(6) 2008, PRODUCT TYPE LEAD; PRODUCT ID 3487A-45, LOT# V061945, IMPLANTED: (B)(6) 2008, PRODUCT TYPE LEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD DIFFICULTY USING THE PATIENT PROGRAMMER DUE TO MEDICAL LIMITATIONS. THE PATIENT HAD MINI STROKES, AND AS A RESULT HAD A HARD TIME REMEMBERING THINGS. IT WAS NOTED THAT THE PATIENT WOULD GET OUT A MANUAL AND START READING IT, BUT AFTER A WHILE HAD TO STOP AND START OVER AGAIN BECAUSE SHE COULDN'T REMEMBER WHAT SHE JUST READ. THE PATIENT EXPERIENCED A SHOCKING OR JOLTING SENSATION. THE PATIENT WAS TURNING ON ALL THE BUTTONS ON THE PATIENT PROGRAMMER BECAUSE SHE COULDN'T FIGURE OUT WHAT TO DO, AND FINALLY GOT THE DEVICE TO TURN OFF. THE PATIENT WAS INSTRUCTED ON HOW TO FIRST TURN DOWN ALL THE DEVICE SETTINGS TO 0.0V USING GROUP ADJUST, THEN TURNING THE STIMULATION ON AND USING GROUP ADJUST TO INCREASE SETTINGS AGAIN. THE PATIENT WAS WALKED THROUGH EACH INDIVIDUAL PROGRAM AND INCREASED OR DECREASED LEVELS BASED ON HOW IT FELT TO THE PATIENT UNTIL THE PATIENT FELT COMFORTABLE. THESE ACTIONS RESOLVED THE REPORTED ISSUE. AT THE END OF THE CALL THE PATIENT WAS HAD STIMULATION BACK ON AND AT THE RIGHT SETTINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ADVANCED STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 37713

Patients

Seq Age Sex Outcome Treatment
1