PTS PANELS CREATININE TEST STRIPS
Report
- Report Number
- 1836135-2012-00001
- Event Type
- Malfunction
- Date Received
- May 2, 2012
- Date of Event
- April 3, 2012
- Report Date
- April 23, 2011
- Manufacturer
- POLYMER TECHNOLOGY SYSTEMS
- Product Code
- JFY
- PMA / PMN Number
- K010456
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
AS PART OF A ROUTINE PRODUCT INVESTIGATION ((B)(4)), A PROGRAMMING ERROR WAS DISCOVERED WITH THE VERSION 2.60 CARDIOCHEK PA METER WHEN USED IN CONJUNCTION WITH CREATININE LOT F103. THE PROGRAMMING ERROR CREATED A POTENTIAL FOR AN ERRONEOUS RESULT DISPLAY FOR ANY VALUES LESS THAN 1.0 MG/DL. THIS ASSAY OF CREATININE ALONG WITH THE CARDIOCHEK IS THE ONLY ASSAY THAT WILL REPORT VALUES LESS THAN 1.0 MG/DL. THE LOT WAS RELEASED IN DECEMBER AND NO COMPLAINTS OF THIS NATURE HAVE BEEN LOGGED AGAINST THIS LOT. THE NORMAL CLINICAL RANGES ARE 0.6-1.2 MG/DL FOR ADULT MALES AND 0.5-1.1 MG/DL FOR ADULT FEMALES. AN ELEVATED CREATININE LEVEL THAT IS OUTSIDE THE NORMAL REFERENCE RANGE IS AN INDICATION OF KIDNEY DAMAGE AND/OR REDUCED KIDNEY FUNCTION AND WOULD REQUIRE IMMEDIATE FOLLOW-UP TESTING. THE PROGRAMMING ERROR ONLY EFFECTS CREATININE RESULTS LESS THAN 1.0 MG/DL AND WOULD NOT BE ASSOCIATED WITH THE ABOVE DESCRIBED CLINICAL USAGES BASED ON THE EFFECTED RANGE OF LESS THAN 1.0 MG/DL. AN ERRONEOUS RECOVERY OF CREATININE IN A RESULT LESS THAN 1.0 MG/DL IS A TOLERABLE TO MANAGEABLE RISK BASED ON THE DEGREE OF ERRONEOUS RECOVERY AND THE FREQUENCY OF OCCURRENCE IF THE ERRONEOUS RECOVERY WILL STILL REPORT THE RESULT AS A VALUE LESS THAN 1.0 MG/DL. IF AN ERRONEOUS RESULT WAS TO OCCUR IN A CREATININE RESULT LESS THAN 1.0 MG/DL, THE SEVERITY WOULD BE ESTIMATED AS A LOW RISK IF THE ERRONEOUS RECOVERY WILL STILL REPORT THE RESULT AS A VALUE LESS THAN 1.0 MG/DL. THE IDENTIFIED PROGRAMMING ERROR ONLY EFFECTS RESULTS LESS THAN 1.0 MG/DL AND ANY ERRONEOUS RESULT PRODUCED WILL REPORT AS A VALUE LESS THAN 1.0 MG/DL. THE FREQUENCY OF THE PRODUCT ERRONEOUS RECOVERY IS ESTIMATED AS LOW TO OCCASIONAL BECAUSE THE ERROR WILL ONLY OCCUR IN RESULTS LESS THAN 1.0 MG/DL. WE HAVE THUS UNDERTAKEN A FIELD CORRECTION FOR THIS LOT WHICH HAS ONLY BEEN DISTRIBUTED TO A VERY LIMITED NUMBER OF CUSTOMERS THROUGH TWO DISTRIBUTORS IN THE USA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PTS PANELS CREATININE TEST STRIPS | CREATININE TEST STRIPS | JFY | POLYMER TECHNOLOGY SYSTEMS | F103 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |