FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT STRIPS

MDR report key: 2556295 · Received April 24, 2012

Report

Report Number
1826988-2012-00197
Event Type
Malfunction
Date Received
April 24, 2012
Date of Event
March 1, 2012
Report Date
March 21, 2012
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
PMA / PMN Number
K111268
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IN SOME COUNTRIES OUTSIDE THE US, CUSTOMER INFORMATION IS NOT PROVIDED DUE TO PRIVACY LAWS.

Description of Event or Problem · 1

GERMAN SALES REPRESENTATIVE RECEIVED CONTROL TESTS OF 180, 176, 188, 148 MG/DL RANGE IS 118-147 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. STRIPS WERE REPLACED AND REPRESENTATIVE IS TO RETURN STRIPS FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR NEXT STRIPS NBW BAYER HEALTHCARE, LLC 2AFEC01B

Patients

Seq Age Sex Outcome Treatment
1 UNK