ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2012-01252
- Event Type
- Malfunction
- Date Received
- May 1, 2012
- Date of Event
- April 6, 2012
- Report Date
- April 11, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS DISPATCHED TO THE SITE ON (B)(6) 2012 FOR THIS EVENT. THE FIELD SERVICE ENGINEER (FSE) VERIFIED INSTRUMENT HARDWARE WITHOUT HAVING TO PERFORM ANY HARDWARE REPAIRS. MDR ASSOCIATED WITH THIS EVENT: 2122870-2012-01252, 2122870-2012-01253, AND 2122870-2012-01254.
THE CUSTOMER REPORTED THAT ERRONEOUS ALPHA-FETOPROTEIN (AFP) RESULTS WERE GENERATED ON THE ACCESS 2 IMMUNOASSAY SYSTEM FOR AN APPROXIMATELY SEVEN PATIENTS ACROSS THREE DAYS. BECKMAN COULTER INC ASSESSMENT OF CUSTOMER SUPPLIED DATA INDICATED THE GENERATION OF IMPRECISE AFP RESULTS ACROSS TWO DAYS. A THIRD DAY WAS IMPLICATED BY THE CUSTOMER AS HAVING GENERATED AN ERRONEOUS AFP RESULT, HOWEVER, ACTUAL PATIENT RESULTS WERE NOT PROVIDED BY THE CUSTOMER FOR THE 3RD DAY. THIS REPORT REPRESENTS THE ERRONEOUS AFP RESULT GENERATED ON AN ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT ON (B)(6) 2012. THE CUSTOMER DID NOT PROVIDE ACTUAL PATIENT RESULTS FOR THIS DATE. UPON REPEAT TESTING ON THE SAME INSTRUMENT, THE REPEAT AFP RESULT WAS REGARDED AS VALID. THE ERRONEOUS AFP RESULT WAS REPORTED OUTSIDE OF THE LABORATORY, HOWEVER, THERE WAS NO DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED OR ATTRIBUTED TO THIS EVENT. ASSAY QUALITY CONTROL RESULTS RECOVERED WITHIN THE CUSTOMER'S ESTABLISHED LIMITS FROM (B)(6) 2012. SYSTEM CHECKS PERFORMED BY THE CUSTOMER PASSED WITHIN INSTRUMENT SPECIFICATIONS ON THE (B)(6) 2012. THE SAMPLE WAS A SERUM SAMPLE WHICH WAS LESS THAN 24 HRS OLD AT THE TIME OF TESTING. THE SAMPLE WAS STORED REFRIGERATED PRIOR TO INITIAL TESTING AND WAS FROZEN, THAWED, MIXED AND RESPUN PRIOR TO REPEAT TESTING. THE REAGENT AND CALIBRATOR LOTS ASSOCIATED WITH THIS EVENT WERE 110505 AND 110341 RESPECTIVELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |