ARCHITECT C8000 SYSTEM
Report
- Report Number
- 1628664-2012-00266
- Event Type
- Malfunction
- Date Received
- May 1, 2012
- Date of Event
- April 10, 2012
- Report Date
- April 11, 2012
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- K980367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATED THAT THE SAMPLES TESTED ON THE I-STAT WERE FROM FINGERSTICKS; WHILE SAMPLES RAN ON THE ARCHITECT C8000 WERE DRAWN FROM A PORT LINE. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED RELATED TO THE CUSTOMER ISSUE. THE ARCHITECT C8000 AND THE I-STAT OPERATIONS MANUALS WERE REVIEWED AND WERE FOUND TO ADEQUATELY ADDRESS THE ISSUE. NO PRODUCT DEFICIENCY OR MALFUNCTION WAS IDENTIFIED.
(B)(4) - (NO CONSEQUENCES OR IMPACT TO PATIENT); (B)(4) - (INCORRECT OR INADEQUATE TEST RESULT). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE ACCOUNT STATED THAT THEY HAVE A PATIENT WHOSE ARCHITECT POTASSIUM RESULTS ARE HIGH COMPARED WITH THE I-STAT. ON (B)(6) 2012 AT 2225: 7.9 MMOL/L ON ARCHITECT VS. 5.1 MMOL/L AT 2000 ON I-STAT; ON (B)(6) 2012 AT 0001: 8.8 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0228: 4.3 MMOL/L ON ARCHITECT VS. 4.1 MMOL/L AT 0316 ON I-STAT; ON (B)(6) 2012 AT 0438: 6.7 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0655: 7.9 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0821: 6.9 MMOL/L ON ARCHITECT VS. 4.0 MMOL/L AT 1010 ON I-STAT; ON (B)(6) 2012 AT NOON: 4.1 MMOL/L ON I-STAT; THE DRAW TIMES WERE NOT THE SAME FOR ARCHITECT VS. I-STAT. SAMPLES FOR LAB WERE ALL GREEN TOP LITHIUM HEPARIN DRAWN THROUGH A PORT WITH SALINE USED TO FLUSH. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT C8000 SYSTEM | JJE | ABBOTT MANUFACTURING INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT ICT MODULE LN 9D28-03| ARCHITECT ICT MODULE LN 9D28-03 |