FDA Adverse Event Malfunction Summary report: N

ARCHITECT C8000 SYSTEM

MDR report key: 2555039 · Received May 1, 2012

Report

Report Number
1628664-2012-00266
Event Type
Malfunction
Date Received
May 1, 2012
Date of Event
April 10, 2012
Report Date
April 11, 2012
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
JJE
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATED THAT THE SAMPLES TESTED ON THE I-STAT WERE FROM FINGERSTICKS; WHILE SAMPLES RAN ON THE ARCHITECT C8000 WERE DRAWN FROM A PORT LINE. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED RELATED TO THE CUSTOMER ISSUE. THE ARCHITECT C8000 AND THE I-STAT OPERATIONS MANUALS WERE REVIEWED AND WERE FOUND TO ADEQUATELY ADDRESS THE ISSUE. NO PRODUCT DEFICIENCY OR MALFUNCTION WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4) - (NO CONSEQUENCES OR IMPACT TO PATIENT); (B)(4) - (INCORRECT OR INADEQUATE TEST RESULT). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT THEY HAVE A PATIENT WHOSE ARCHITECT POTASSIUM RESULTS ARE HIGH COMPARED WITH THE I-STAT. ON (B)(6) 2012 AT 2225: 7.9 MMOL/L ON ARCHITECT VS. 5.1 MMOL/L AT 2000 ON I-STAT; ON (B)(6) 2012 AT 0001: 8.8 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0228: 4.3 MMOL/L ON ARCHITECT VS. 4.1 MMOL/L AT 0316 ON I-STAT; ON (B)(6) 2012 AT 0438: 6.7 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0655: 7.9 MMOL/L ON ARCHITECT; ON (B)(6) 2012 AT 0821: 6.9 MMOL/L ON ARCHITECT VS. 4.0 MMOL/L AT 1010 ON I-STAT; ON (B)(6) 2012 AT NOON: 4.1 MMOL/L ON I-STAT; THE DRAW TIMES WERE NOT THE SAME FOR ARCHITECT VS. I-STAT. SAMPLES FOR LAB WERE ALL GREEN TOP LITHIUM HEPARIN DRAWN THROUGH A PORT WITH SALINE USED TO FLUSH. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT C8000 SYSTEM JJE ABBOTT MANUFACTURING INC

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT ICT MODULE LN 9D28-03| ARCHITECT ICT MODULE LN 9D28-03