FDA Adverse Event Malfunction Summary report: N

VERSO AIRWAY ACCESS ADAPTER

MDR report key: 2554635 · Received May 6, 2010

Report

Report Number
2050001-2010-00003
Event Type
Malfunction
Date Received
May 6, 2010
Date of Event
March 10, 2010
Report Date
May 5, 2010
Manufacturer
CAREFUSION
Product Code
BZA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SAMPLE WAS NOT RECEIVED FOR EVAL; THEREFORE, WE ARE UNABLE TO DETERMINE A ROOT CAUSE FOR THE ISSUE REPORTED. THIS IS THE FIRST FORMAL REPORT OF BREAKAGE INVOLVING BREAKAGE WITH SPLINTERING OF THE PLASTIC IN SEPARATE PIECES. WE WILL CONTINUE TO MONITOR OUR PRODUCTS PERFORMANCE FOR ADDITIONAL REPORTS OF THIS TYPE OF BREAKAGE. WE DO NOT ANTICIPATE PHYSICAL BREAKAGE OF THE PLASTIC IN THE MANNER DESCRIBED IN THE REPORT WITHOUT CONSIDERABLE FORCE BEING APPLIED TO THE COMPONENT.

Description of Event or Problem · 1

PT IN THE ICU AND WHILE THE PT WAS BEING TURNED, THE NURSE HEARD AIR LEAK. THE VERSO ADAPTER HAD SNAPPED IN TWO AND PART OF IT REMAINED IN THE PT. THE PT BECAME DISCONNECTED FROM THE VENTILATOR. ONE NURSE HELD ONTO THE ADAPTER AND KEEPING IN-LINE SUCTION TOGETHER WHILE THE OTHER WAS USING THE AMBU BAG TO VENTILATE THE PT. PIECE WAS REMOVED FROM PT'S THROAT WITH FORCEPS WITHOUT INJURY. IN-LINE SUCTION WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSO AIRWAY ACCESS ADAPTER VERSO AIRWAY ACCESS ADAPTER BZA CAREFUSION CSC100 09060014

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other