ARCHITECT CYCLOSPORINE
Report
- Report Number
- 1415939-2012-00178
- Event Type
- Malfunction
- Date Received
- April 30, 2012
- Report Date
- April 18, 2012
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- MKW
- PMA / PMN Number
- K080751
- Removal / Correction Number
- 1415939-10/10/11-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE CAUSE OF THE ARCHITECT CYCLOSPORINE IMPRECISION IS DUE TO AN INTERACTION BETWEEN THE ARCHITECT CYCLOSPORINE ASSAY AND THE RESIN USED TO PRODUCE CERTAIN ARCHITECT REACTION VESSEL (RV) LOTS. ASSAY PRECISION CAN BE IMPACTED IF RVS CONTAINING DIFFERENT RESIN ARE MIXED IN THE ARCHITECT HOPPER AND USED TO TEST ARCHITECT CYCLOSPORINE. ALL ARCHITECT CYCLOSPORINE CUSTOMERS WILL BE INSTRUCTED TO ONLY RUN ARCHITECT CYCLOSPORINE USING RVS MANUFACTURED WITH RESIN FROM THE SAME SUPPLIER. RV LOTS BEGINNING WITH LOT 06083P100 AND HIGHER CAN BE USED TOGETHER. RV LOTS LOWER THAN 06083P100 CAN BE USED TOGETHER.
THE CUSTOMER STATED CYCLOSPORINE QUALITY CONTROLS WERE OUT OF RANGE HIGH ON AN ARCHITECT (B)(4) ANALYZER. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.