DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2012-00184
- Event Type
- Injury
- Date Received
- April 29, 2012
- Date of Event
- March 28, 2012
- Report Date
- January 10, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4).
THE SYSTEM LOGS WERE REVIEWED BY AN ISI TECHNICAL SUPPORT ENGINEER AND NO ERRORS WERE OBSERVED WHICH INDICATE THAT THERE WERE ISSUES WITH THE PSM ARMS. THE PSMS ARE INSTRUMENT ARMS LOCATED ON THE PATIENT CART THAT PROVIDE THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE ROOT CAUSE OF THE SURGICAL COMPLICATION EXPERIENCED BY THE PATIENT IS CURRENTLY UNDETERMINED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO GAIN ADDITIONAL INFORMATION ABOUT THE EVENT, BUT THE CUSTOMER HAS YET TO PROVIDE FURTHER DETAILS. IF ADDITIONAL EVENT DETAILS BECOME AVAILABLE A FOLLOW UP MDV WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING A DA VINCI SI GASTRIC BYPASS PROCEDURE THE SURGEON WAS UNABLE TO MOVE PATIENT SIDE MANIPULATOR (PSM) ARM 2 FROM THE SURGEON CONSOLE. AFTER RESTARTING THE SYSTEM, PSM ARM 2 WORKED FINE. AFTER 15 MINUTES OF OPERATION THE SURGEON LOST CONTROL OF PSM ARM 1, RESULTING IN DAMAGE TO THE PATIENT'S STOMACH. NO FURTHER INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.0P6 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES |