CARESTREAM DRX-EVOLUTION SYSTEM /FF WAL STAND
Report
- Report Number
- 1317307-2012-00001
- Event Type
- Malfunction
- Date Received
- April 19, 2012
- Date of Event
- March 14, 2012
- Report Date
- April 19, 2012
- Manufacturer
- CARESTREAM HEALTH, INC.
- Product Code
- KPR
- PMA / PMN Number
- K091889
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE POSITIONING OF THE WALL STAND USING MOTORIZED MOVEMENT SHOULD NOT BE COMPLETED WITH THE PT STANDING BY THE BUCKY OR LEANING ON THE LATERAL ARM SUPPORT OF THE BUCKY ACCORDING TO THE DRX EVOLUTION USER INSTRUCTIONS. CAUTION STATEMENTS IN THE USER GUIDE #9G5640 STATE: "WITH AUTO TRACKING ENABLED THE AUTOMATIC MOVEMENT APPROXIMATELY FOLLOWS THE OTC AND THE BUCKY. NO INDIVIDUALS OR OBJECTS SHOULD BE NEAR OR ON THE WALL STAND. IMPEDING THE MOTION OF THE OTC OR ADDING WEIGHT TO THE WALL STAND CAN CAUSE DAMAGE. OBSERVE THE PT AT ALL TIMES ESPECIALLY WHEN USING AUTOMATIC MOTION. THE ARM SUPPORT IS NOT DESIGNED TO BEAR A PT'S WEIGHT. IT IS INTENDED TO BE USED AS AN OVERHEAD HAND GRIP, POSITIONING THE ARMS OF THE PT WHILE THE X-RAY IS TAKEN. MAKE SURE THE EQUIPMENT DOES NOT COME INTO CONTACT WITH THE PT WHEN USING AUTOMATIC MOTION." SERVICE ENGINEERING REPAIRED THE ISSUE AT THE CUSTOMER SITE AND THERE HAVE NOT BEEN ANY FURTHER ISSUES WITH THE OPERATION OF THE EQUIPMENT. A CONFIRMED ROOT CAUSE HAS NOT BEEN IDENTIFIED, HOWEVER BASED UPON THE INVESTIGATION IT IS BELIEVED THAT THE RJ45 TYPE 1 CONNECTOR ON THE ONE END OF THE COMMUNICATION CABLE MAY HAVE HAD CRIMPED WIRES WITHIN THE CONNECTOR LEADING TO INTERMITTENT COMMUNICATION PROBLEMS. THE TMC BOARD, LEFT KEYPAD, AND LEFT AND RIGHT CABLES WERE REPLACED ON THE WALL STAND.
CUSTOMER SITE REPORTS 2 ISSUES THAT RELATE TO DOWNWARD MOVEMENT OF THE WALL STAND BUCKY. THERE WERE NO PT INJURIES OR ADVERSE EVENTS AS A RESULT OF THIS ISSUE. EACH ISSUE RELATED TO THE RADIOLOGY TECHNOLOGIST (RT) PREPARING TO COMPLETE A CHEST X-RAY AND PRESSED THE AUTO TRACKING BUTTON WHILE THE PT WAS EITHER LEANING OR HOLDING ONTO THE ARMS OF THE WALL STAND BUCKY. WHEN PRESSING THE DOWN BUTTON ON THE LEFT KEYPAD THE BUCKY BEGAN TO MOVE DOWNWARD STOPPING 2 INCHES FROM THE FLOOR. ONCE THE BUCKY DID STOP MOVING, THE WALL STAND MOVED BACK UP, MANUALLY IN ONE INSTANCE BECAUSE THE BUTTONS WOULD NOT WORK. ONCE THE RT PRESSED THE AUTO TRACKING BUTTON AGAIN THERE WERE NO OTHER MALFUNCTIONS FOR THE REMAINDER OF THE DAY AND THE RT CONTINUED TO USE THE EQUIPMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARESTREAM DRX-EVOLUTION SYSTEM /FF WAL STAND | DRX-EVOLUTION WALL STAND | KPR | CARESTREAM HEALTH, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |