FDA Adverse Event Malfunction Summary report: N

CAROLON

MDR report key: 254962 · Received December 14, 1999

Report

Report Number
254962
Event Type
Malfunction
Date Received
December 14, 1999
Date of Event
November 4, 1999
Report Date
November 29, 1999
Manufacturer
CAROLON CO.
Product Code
JOW
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

ON 11/4/99, A CAROLON TRINITY COMPRESSION PUMP, IN USE ON PT, WAS FOUND BY NURSE TO BE SMOKING. THE PUMP WAS UNPLUGGED AND REMOVED FROM SERVICE. THE PUMP WAS EXAMINED BY CLINICAL ENGINEERING SERVICE AND IT WAS FOUND THAT THE PIN FOR THE HOT SIDE OF THE MALE POWER RECEPTACLE ON THE UNIT WAS PUSHED IN AND BURNT AND THE CASING AROUND IT WAS CRACKED. A HOLE WAS BURNT THROUGH THE REMOVABLE POWER CORD WHERE IT CONNECTED TO THE UNIT. ON 11/23/99, A GROUP OF 6 CAROLON TRINITY COMPRESSION PUMPS, WHICH WERE REPORTED AS BROKEN, WERE EXAMINED BY CLINICAL ENGINEERING BEFORE BEING SENT BACK TO THE MFR FOR REPLACEMENT UNDER THE LEASE CONTRACT. IT WAS FOUND THAT ALL 6 PUMPS ALSO HAD THE PIN PUSHED IN AND CRACKS IN THE RECEPTACLE CASING. THE REMAINING CAROLON TRINITY COMPRESSION PUMPS WERE REMOVED FROM SERVICE ASAP AFTER FINDING THIS PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAROLON COMPRESSION PUMP JOW CAROLON CO. TRINITY DVT SYSTEM *

Patients

Seq Age Sex Outcome Treatment
1 * Other