FDA Adverse Event Injury Summary report: N

PILLCAM PATENCY CAPSULE

MDR report key: 2549498 · Received April 23, 2012

Report

Report Number
9710107-2012-01373
Event Type
Injury
Date Received
April 23, 2012
Report Date
April 23, 2012
Manufacturer
GIVEN IMAGING LTD.
Product Code
NEZ
PMA / PMN Number
K053639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AGILE CAPSULE IS AN ACCESSORY TO THE PILLCAM VIDEO CAPSULE AND ITS INTENDED USE IS TO VERIFY ADEQUATE PATENCY OF THE GASTROINTESTINAL TRACT PRIOR TO ADMINISTRATION OF THE PILLCAM VIDEO CAPSULE IN PATIENTS WITH KNOWN OR SUSPECTED STRICTURES. THE PATIENT HAD A HISTORY OF SURGERY AND EXPECTED POST SURGICAL ADHESIONS. THE CAPSULE WAS PROBABLY RETAINED DUE TO ADHESIONS.

Description of Event or Problem · 1

A CUSTOMER REPORTED OF A PATIENT THAT HAD A 42 HOUR RETENTION OF A PILLCAM PATENCY CAPSULE AND NEEDED MEDICAL INTERVENTION. THE PATIENT, WHO HAS A HISTORY OF SURGERY AND EXPECTED POST-SURGICAL ADHESIONS, WAS HOSPITALIZED AFTER COMPLAINING OF PAIN, NAUSEA AND VOMITING. ACCORDING TO THE CUSTOMER, THE PATIENT HAD INTERMITTENT OBSTRUCTIVE SYNDROME, PROBABLY FROM ADHESIONS. A CT SCAN SHOWED THE CAPSULE IS INTACT AND THE PATIENT EVENTUALLY WAS SENT TO SURGERY. THE CAPSULE WAS INTACT IN THE RESECTED BOWEL SEGMENT. IT IS POSSIBLE THAT THE PATIENT ALSO HAD CROHN'S DISEASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PILLCAM PATENCY CAPSULE PILLCAM PATENCY CAPSULE NEZ GIVEN IMAGING LTD. PILLCAM PATENCY CAP NA

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R