TRIAGE CARDIAC PANEL 25 TEST
Report
- Report Number
- 2027969-2012-00556
- Event Type
- Death
- Date Received
- April 19, 2012
- Date of Event
- March 19, 2012
- Report Date
- April 19, 2012
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- NBC
- PMA / PMN Number
- K030286
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AS OF (B)(6) 2012, THERE ARE A TOTAL OF 6 COMPLAINTS AGAINST DEVICE LOT W49719, 4 OF THE COMPLAINTS INVOLVING TNI RECOVERY. PRODUCT SUPPORT HAS PREVIOUSLY CONDUCTED TESTING OF W49719 FOR THREE PREVIOUS COMPLAINTS. THE DEVICE LOT APPEARED TO FUNCTION WITHIN THE MFR'S RELEASE SPECIFICATIONS WITH THE POSITIVE AND NEGATIVE CONTROL SAMPLES. NO PRODUCT DEFICIENCY WAS ESTABLISHED. CUSTOMER PROVIDED CKMB AND MYOGLOBIN RESULTS ALONG WITH ELEVATED TNI RESULTS. CKMB SUPPORTS THE TNI VALUES. NO SAMPLE WAS RETURNED FOR TESTING. (B)(4) WAS OPENED TO ADDRESS ELEVATED TNI AT LOW END CONCENTRATIONS. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME.
CUSTOMER REPORTED POTENTIAL FALSE POSITIVE TROPONIN I (TNI) RESULTS ON THE TRIAGE CARDIAC PANEL TEST VS. SIEMENS VISTA ANALYZER. A SAMPLE WAS DRAWN FROM A (B)(6) FEMALE PT AND TESTED USING THE TRIAGE CARDIAC PANEL TEST. THE TROPONIN I RESULTS WERE ABOVE THEIR CUT-OFF. TWO HRS LATER, THE PT'S TNI VALUE WAS HIGHER. THE FOLLOWING DAY TESTING WAS PERFORMED USING THE SIEMENS VISTA WITH A NEGATIVE RESULT. PT'S INITIAL DIAGNOSIS WAS: CVA STROKE. NO FINAL DIAGNOSIS PROVIDED. A MESSAGE RECEIVED FROM AN ALERE ACCOUNT EXECUTIVE ON (B)(6) 2012 STATED THAT THE 'CUSTOMER BELIEVES THAT THE PT HAS EXPIRED'. NO FINAL DIAGNOSIS, CAUSE OF DEATH OR TIME/DATE OF DEATH PROVIDED. THERE IS NO INDICATION THAT THE PT'S DEATH WAS RELATED TO THE RESULTS OBTAINED FROM THE TRIAGE CARDIAC PANEL TEST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIAGE CARDIAC PANEL 25 TEST | CARDIAC MARKER TEST | NBC | ALERE SAN DIEGO, INC | 97000HS | W49719B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 91 YR | Death |