FDA Adverse Event Malfunction Summary report: N

BGSTAR BGMS

MDR report key: 2548494 · Received April 17, 2012

Report

Report Number
3004637226-2012-00010
Event Type
Malfunction
Date Received
April 17, 2012
Date of Event
December 15, 2011
Report Date
April 16, 2012
Manufacturer
AGAMATRIX, INC.
Product Code
NBW
PMA / PMN Number
K072413
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METER S/N (B)(4) AND TEST STRIP LOT # IJ14WC13C, EXP: 11.2012 WITH RANGE 6.0 - 9.0 MMOL/L WERE TESTED ON (B)(6) 2012. PERFORMANCE RESULTS AS FOLLOWS: 8.6 MMOL/L, 8.5 MMOL/L, AND 8.5 MMOL/L. THE COMPLAINT COULD NOT BE CONFIRMED. THE FUNCTION TEST SHOWED THE BG STAR METER WORKED PROPERLY AND MET SPECIFICATIONS.

Description of Event or Problem · 1

AN EVENT OCCURRED WITH THE BGSTAR SYSTEM WHERE INDIRECT HARM OCCURRED BECAUSE THE PATIENT INJECTED A HIGHER INSULIN DOSE AND WENT INTO HYPOGLYCEMIA BASED ON READINGS ON THE BGSTAR WHICH WERE PRESUMED TO BE TOO HIGH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BGSTAR BGMS BLOOD GLUCOSE METER NBW AGAMATRIX, INC. 8000-05000

Patients

Seq Age Sex Outcome Treatment
1 Other