REAGENT RED BLOOD CELLS BIOTESTCELL 3
Report
- Report Number
- 9610824-2012-00064
- Event Type
- Malfunction
- Date Received
- April 25, 2012
- Date of Event
- March 28, 2012
- Report Date
- April 25, 2012
- Manufacturer
- BIO-RAD MEDICAL DIAGNOSTICS GMBH
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS IS OUR COMBINED INITIAL AND FINAL REPORT ON THIS INCIDENT
THE CUSTOMER REPORTED A FALSE NEGATIVE ANTIBODY SCREENING TEST OF A PATIENT WITH A KNOWN ANTI-E AND ANTI-K ON TANGO. THE IDENTIFICATION OF THE ANTIBODIES WAS PERFORMED USING THE ORTHO GEL SYSTEM. THE PATIENT SAMPLE THAT HAD CAUSED THE ALLEGEDLY FALSE NEGATIVE TEST RESULT WAS SENT TO US FOR QUALITY CONTROL, BUT NONE OF THE SUPPOSEDLY DEFECTIVE PRODUCT WAS RETURNED. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THE RETAINED SAMPLE OF BIOTESTCELL 3 WITH THE PROVIDED PATIENT SAMPLE. THE SAMPLE REACTED NEGATIVELY. IN FURTHER TESTINGS THE PATIENT SAMPLE SHOWED A POSITIVE DIRECT ANTIGLOBULIN TEST (DAT) AND A VERY WEAK POSITIVE REACTION (+/- ) IN THE GEL METHOD. THE PATIENT MATERIAL WAS NOT SUFFICIENT TO CARRY OUT FURTHER TESTINGS. THE RETAINED SAMPLE OF BIOTESTCELL 3 WAS TESTED WITH DIFFERENT ANTIBODIES, INCLUSIVE AN ANTI-K AND ANTI-E. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTIONS. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THE ALLEGEDLY DEFECTIVE LOT OF BIOTESTCELL 3 FUNCTIONS CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. A CHECK-UP OF THE AFFECTED TANGO OPTIMO GAVE NO INSTIGATION TO SUSPECT THE PERFORMANCE OF THE DEVICE TO BE PART OF THE REPORTED PROBLEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REAGENT RED BLOOD CELLS BIOTESTCELL 3 | BIOTESTCELL 3 | KSZ | BIO-RAD MEDICAL DIAGNOSTICS GMBH | 8207011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SOLIDSCREEN II STRIP, LOT 8137050| ERYTYPECELL A1,B, LOT 8207011| ANTI-HUMAN-GLOBULIN, LOT 812112004| ERYTYPE S, LOT 8123010| BROMELIN, LOT 815105SS| TANGO OPTIMO, NO. (B)(4)| MLB2, LOT 8114021 |