SAFECLINITUBES
Report
- Report Number
- 3002807968-2012-00006
- Event Type
- Malfunction
- Date Received
- April 20, 2012
- Date of Event
- February 28, 2012
- Report Date
- March 27, 2012
- Manufacturer
- RADIOMETER MEDICAL APS.
- Product Code
- GIO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
ON THIS NEONATAL DEPARTMENT THEY HAVE HAD TWO DIFFERENT TYPES OF PLASTIC CAPILLARY TUBES, I.E. SAFECLINITUBES D941P-NA-240-85 X 3, PART NO. 942-891, AND SAFECLINITUBES D957P-70-100 X 1, PART NO. 942-892. THESE TWO TYPES OF PLASTIC CAPILLARIES WERE MIXED UP BY THE USERS (NURSES AND PHYSICIANS) WHEN BLOOD SAMPLE MEASUREMENTS HAVE BEEN PERFORMED. THEY HAVE USED THE CAPILLARY TYPE WHICH CONTAINS SODIUM HEPARIN WHEN THEY PERFORMED BLOOD SAMPLE MEASUREMENTS TO GET ELECTROLYTE RESULTS. THIS IS A USE ERROR. WE HAVE REVIEWED THE PACKAGE LABELING AND INSERT FOR THE SAFECLINITUBES (ATTACHED). BOTH THE PACKAGE AND THE INSERT STATES WHICH TYPE OF HEPARIN THE TUBES CONTAIN. FURTHERMORE THE ABL835 ANALYZER OPERATORS MANUAL CONTAINS THE FOLLOWING CAUTION (P. 1-5 'LIMITATIONS OF USE AND KNOWN INTERFERING SUBSTANCES,'): 'ANTICOAGULANTS THAT CONTAIN SODIUM SALTS WILL GIVE ERRONEOUSLY HIGH CNA+ RESULTS. SODIUM FLUORIDE WITH OR WITHOUT EDTA AND OXALATE (DI NA) INFLUENCE CGLU RESULTS. SODIUM FLUORIDE GIVES ERRONEOUSLY HIGH CNA+ AND LOW CCA2+, CGLU AND CLAC RESULTS. TRI SODIUM CITRATE INFLUENCES CNA+, CK+ AND CGLU RESULTS.'
AN ABL835 ANALYZER REPORTED RESULTS OF BLOOD SAMPLE MEASUREMENTS OF NEONATAL PTS WITH AN INCREASED CA VALUE OF APP. 0, 3 MMOL/L HIGHER THAN EXPECTED WHEN MEASUREMENTS WERE PERFORMED ON CAPILLARY BLOOD SAMPLES USING RADIOMETER PLASTIC CAPILLARY TUBES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SAFECLINITUBES | PLASTIC CAPILLARY TUBES | GIO | RADIOMETER MEDICAL APS. | D941P-NA-240-85 X 3 | DW 01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |