ZIMMER M/L TAPER MODULAR FEMORAL STEM
Report
- Report Number
- 1822565-2012-00881
- Event Type
- Injury
- Date Received
- April 20, 2012
- Date of Event
- March 20, 2012
- Report Date
- March 23, 2012
- Manufacturer
- ZIMMER INC
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: IT IS REPORTED THAT THE PT EXPERIENCED A MEDIAL FRACTURE OF THE PROXIMAL FEMUR AND WAS REVISED. SURGICAL NOTES WERE NOT PROVIDED. X-RAYS WERE NOT PROVIDED; IT IS UNK WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. IN GENERAL, PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS INCLUDE: AGE, BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT), AND RELEVANT MEDICAL HISTORY. ADHERENCE TO REHABILITATION PROTOCOL IS UNK. A DEFINITIVE CONCLUSION CANNOT BE REACHED WITH THE INFO PROVIDED. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THEY EXPERIENCED A "POPPING" SOUND. X-RAYS REVEALED A MEDIAL FRACTURE OF THE PROXIMAL FEMUR. THE PT WAS REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZIMMER M/L TAPER MODULAR FEMORAL STEM | KWA | ZIMMER INC | 61796983 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |