MAQUET SAS
Report
- Report Number
- 9710055-2012-00006
- Event Type
- Malfunction
- Date Received
- March 28, 2012
- Date of Event
- February 24, 2012
- Report Date
- February 24, 2012
- Manufacturer
- MAQUET S.A.
- Product Code
- FTD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
THE HOSPITAL STAFF REPORTED TO A MAQUET FIELD SERVICE TECH (FST) ON SITE THAT THE FASTENING HARDWARE (12 SCREWS) HAD SHEARED OFF. HOWEVER, THE HOSPITAL STAFF DID NOT PROVIDE ANY EXPLANATION AS TO HOW ALL 12 SCREWS/BOLTS FAILED AT THE SAME TIME. THE FST WAS INFORMED THAT ALL OF THE HARDWARE WAS THROWN OUT BY THE HOSPITAL STAFF AND ANY REMNANTS OF THE SCREWS IN THE BOOM WERE DRILLED OUT TO FACILITATE A REPAIR. THE HOSPITAL FAILED TO CAPTURE ANY PHOTOS OF THE BROKEN PARTS BEFORE DISCARDING THEM. THE HANAUPORT SERIES OPERATING MANUAL MENTIONS THAT PENDANT SYSTEMS ARE TO BE SERVICED BY MAQUET OR AN AUTHORIZED CUSTOMER SERVICE ENGINEER ONCE A YEAR. INCLUDED IN THE YEARLY MAINTENANCE PROGRAM ARE SEVERAL MECHANICAL SAFETY VERIFICATIONS. THE HOSPITAL IS THE PRIMARY SERVICE PROVIDER FOR THIS UNIT. IT DID NOT PROVIDE MAQUET ANY SERVICE RECORDS FOR THIS DEVICE. THIS UNIT HAS PRESENTLY BEEN REMOVED FROM SERVICE PENDING THE DELIVERY OF NEW HARDWARE. THIS OCCURRENCE IS THE ONLY ONE REPORTED TO MAQUET, OUT OF (B)(4) UNITS SOLD WORLDWIDE. EXEMPTION# (B)(4). MAQUET MEDICAL SYSTEMS USA SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MFG FACILITY. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.
CUSTOMER REPORTED THAT AN EQUIPMENT BOOM BROKE AND CAME APART IN OPERATING ROOM 4. NO INJURIES WERE REPORTED. THERE WAS NO PT IN THE ROOM AT THE TIME OF THE INCIDENT. IT HAPPENED AS THE ROOM WAS BEING TURNED OVER. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAQUET SAS | FTD | MAQUET S.A. | HANAUPORT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |