PLASMABLADE TISSUE DISSECTION DEVICE
Report
- Report Number
- 3007069406-2012-00083
- Event Type
- Other
- Date Received
- April 5, 2012
- Date of Event
- March 14, 2012
- Report Date
- March 14, 2012
- Manufacturer
- MEDTRONIC ADVANCED ENERGY LLC
- Product Code
- GEI
- PMA / PMN Number
- K083415
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
VISUAL INSPECTION OF THE RETURNED DEVICE NOTED A WHITE ASH AT THE DISTAL TIP OF THE HOUSING. THE HOUSING APPEARED TO BE BURNT AS PER THE REPORTED EVENT. A REVIEW OF THE LOT HISTORY RECORD (LHR) FOR THIS LOT NUMBER FOUND THAT THE DEVICE MET ASSEMBLY AND PRODUCT SPECIFICATIONS, NO ANOMALIES WERE FOUND DURING MFG. BENCH TOP INVESTIGATION FOUND THAT THE SILICONE MATERIAL IN THE TONSIL TIP CAN WITHSTAND THE HIGH TEMPS SEEN DURING NORMAL USE. WHEN EXPOSED TO TEMPS IN EXCESS OF 400 DEGREES CELSIUS, IT WILL BURN, LEAVING A WHITE ASH RESIDUE BUT WILL NOT GENERATE ANY FLAME. IT WAS DETERMINED THAT IN A HIGH OXYGEN ATMOSPHERE, AN ACTIVE TONSIL TIP CAN IGNITE THE OXYGEN AND BURN THE TONSIL TIP HOUSING. IT WAS NOT POSSIBLE TO REPRODUCE THE BURNT TONSIL TIP HOUSING UNDER NORMAL ATMOSPHERIC CONDITIONS. END OF REPORT.
IT WAS REPORTED THAT THE TONSIL TIP IGNITED. NO PT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLASMABLADE TISSUE DISSECTION DEVICE | GEI | MEDTRONIC ADVANCED ENERGY LLC | PLASMABLADE TNA | 51362 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | NOT AVAILABLE |