ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2012-01155
- Event Type
- Malfunction
- Date Received
- April 19, 2012
- Date of Event
- February 28, 2012
- Report Date
- March 21, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
NO SAMPLE COLLECTION OR CENTRIFUGATION DATA WAS SUPPLIED TO DATE. SYSTEM CHECK DATA SUPPLIED BY THE CUSTOMER INDICATED ALL PORTIONS PASSED WITHIN PUBLISHED SPECIFICATIONS. UPON INVESTIGATION OF THE CUSTOMER SUPPLIED DATA INDICATES THAT THE BACKGROUND RELATIVE LIGHT UNITS (RLUS) BEGIN INCREASING AFTER A BHCG REAGENT PACK IS PUT INTO USE ON THE SYSTEM. IN COMPARING THE CUSTOMER'S DATA TO DATA FROM AN UNAFFECTED REAGENT PACK SHOWS THAT THERE IS A STEADY INCREASE IN RLU AND DOSE RECOVERY AS TESTING THAT IS PERFORMED WITH AN "AFFECTED" PACK. THIS INCREASE IN RLU MANIFESTS ITSELF AS ELEVATED L1 QC AND LOW BHCG PATIENTS' SAMPLES. IT WAS NOTED IN THE CUSTOMER DATA THAT WHEN THE CUSTOMER CHANGES TO A FRESH REAGENT PACK, THE RLUS RETURN TO THE ORIGINAL BACKGROUND LEVEL. THE DATA SHOWS THAT WHEN THE CUSTOMER CALIBRATES USING AN "AFFECTED" PACK WITH ELEVATED RLUS, LEVEL 1 QC WILL RECOVER WITHIN RANGE AS THE CALIBRATION CURVE HAS CALCULATED THE RLU INCREASE. HOWEVER, THE SYSTEM ALLOWS MULTIPLE REAGENT PACKS TO BE LOADED FOR A SINGLE ASSAY AND THE SYSTEM CAN AUTOMATICALLY SWITCH TO A NEW "UNAFFECTED" REAGENT PACK ONCE THE AFFECTED PACK IS EMPTY. IN THIS SITUATION, WHEN THE SYSTEM CHANGES TO A FRESH PACK WITH NORMAL BACKGROUND RLU RECOVERY, THE CALCULATED DOSE FOR SAMPLES TESTED WITH THE FRESH PACK WILL BE LOWER DUE TO THE BIAS IN THE CALIBRATION CURVE GENERATED FROM THE "AFFECTED" PACK. THE CUSTOMER CAN POTENTIALLY OBTAIN FALSE NEGATIVE RESULTS FOR PATIENTS' SAMPLES ON THE LOWER END OF THE CALIBRATION CURVE IN SUCH A SITUATION. THE CUSTOMER IS NOT REQUIRED TO QC EACH NEW REAGENT PACK BEFORE USE. THEREFORE THIS CHANGE IN DOSE RESPONSE IS UNSEEN TO THE CUSTOMER UNTIL THE NEXT TIME THEY RUN QC. THERE IS NO INDICATION THAT A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT TO DATE; HOWEVER A SERVICE REQUEST IS SCHEDULED. A DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE. THE FOLLOWING MDRS (TOTAL COUNT: 7) LISTED BELOW INCLUDES THE COMPLETE LIST OF REPORTS SUBMITTED FOR THIS EVENT THAT OCCURRED AT THIS CUSTOMER SITE: 2122870-2012-01147, 2122870-2012-01155, 2122870-2012-01156, 2122870-2012-01157, 2122870-2012-01158, 2122870-2012-01159, 2122870-2012-01160.
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) IN REGARDS TO OBTAINING ELEVATED BETA - HUMAN CHORIONIC GONADOTROPIN (B-HCG) QC LEVEL 1 RESULTS ON THE ACCESS 2 IMMUNOASSAY SYSTEM USED IN CONJUNCTION WITH TWO (2) DIFFERENT ACCESS TOTAL B-HCG REAGENTS (LOT # 170151 AND 170170). THE ERRONEOUS QC RESULTS WERE GENERATED OVER THE COURSE OF SEVERAL DAYS RANGING FROM (B)(6) 2012. THIS REPORT DOCUMENTS THE ELEVATED QC RESULTS OBTAINED ON (B)(6) 2012 USING TOTAL B-HCG REAGENT LOT # 170151. BHCG LEVEL 1 QC RECOVERED GREATER THAN 2 SD HIGH. LEVEL 1 WAS REPEATED AND RECOVERED THE SAME GREATER THAN 2 SD HIGH QC RESULTS. THE CUSTOMER LOADED A NEW REAGENT FROM THE SAME REAGENT LOT, PERFORMED QC, AND LEVEL 1 WAS ONCE AGAIN WITHIN ACCEPTABLE 2 SD RANGE. NO PATIENT RESULTS WERE REPORTED TO HAVE BEEN AFFECTED AS A RESULT OF THIS EVENT. HOWEVER, THE POTENTIAL FOR ERRONEOUS YET BELIEVABLE PATIENTS' RESULTS DOES EXIST UPON RECUR. THESE RESULTS MAY CAUSE INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | ACCESS® 2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |