FDA Adverse Event Injury Summary report: N

OT VERIO IQ METER

MDR report key: 2542121 · Received April 19, 2012

Report

Report Number
3008382007-2012-01558
Event Type
Injury
Date Received
April 19, 2012
Date of Event
April 11, 2012
Report Date
April 12, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. THE 510 (K) # IS K110637.

Description of Event or Problem · 1

ON (B)(6), 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH VERIO IQ METER WAS TESTING IN THE SETTINGS MODE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE CCA WAS ADVISED BY THE PATIENT THAT THE ALLEGED ISSUE BEGAN ON (B)(6), 2012 AT 8:30AM. THE PATIENT STATED THAT SHE DOES NOT TAKE ANY MEDICATION TO MANAGE HER DIABETES AND DENIED MAKING ANY CHANGES TO HER USUAL DIABETES MANAGEMENT ROUTINE IN RESPONSE TO THE REPORTED ISSUE. ONE HOUR AFTER THE ALLEGED ISSUE OCCURRED, THE PATIENT CLAIMED TO HAVE DEVELOPED SYMPTOMS OF "COLD SWEATS AND HUNGER" BUT DENIED RECEIVING ANY TREATMENT IN RESPONSE TO THE SYMPTOMS. DURING THE TROUBLESHOOTING, THE CCA WAS ABLE TO WALK THE PATIENT THROUGH PROPER USAGE OF THE SUBJECT METER PER THE OWNER'S BOOKLET AND THE REPORTED ISSUE WAS RESOLVED. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. ALTHOUGH THE PATIENT DENIED RECEIVING ANY TREATMENT AFTER THE ALLEGED ISSUE BEGAN, THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3246694

Patients

Seq Age Sex Outcome Treatment
1 65 YR Life Threatening