FDA Adverse Event Malfunction Summary report: N

ANGLED DELIVERY DEVICE, GREENLIGHT XPS

MDR report key: 2541024 · Received April 12, 2012

Report

Report Number
2937094-2012-00347
Event Type
Malfunction
Date Received
April 12, 2012
Date of Event
July 6, 2011
Report Date
July 7, 2011
Manufacturer
AMERICAN MEDICAL SYSTEMS, INNOVATION CENTER - SILICON VALLEY
Product Code
GEX
PMA / PMN Number
K100746
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AND ANALYZED. THE CUSTOMERS INITIAL REPORT INDICATED A CONTINUOUS FIBERLIFE MESSAGE, WHICH IS NOT CONSIDERED A REPORTABLE ISSUE. UPON ANALYSIS OF THE RETURNED FIBER, THE FIBER WAS FOUND TO HAVE A CIRCUMFERENTIAL FRACTURE PROXIMAL TO THE FIBER/CAP FUSION ZONE. THE CAP WAS ALSO FOUND TO HAVE BURNT ON DETRITUS AND DEVITRIFICATION OF THE CAP OUTPUT WINDOW. THE FIBER CONDITION WOULD RESULT IN ACTIVATION OF THE SYSTEMS FIBERLIFE FUNCTION WHICH WILL MODULATE (PULSE) THE OUTPUT BEAM AND SEND THE SYSTEM TO STANDBY MODE. BASED ON THE ANALYSIS FINDINGS, THE CIRCUMFERENTIAL FRACTURE CONDITION MAY RESULT IN A FORWARD FIRING CONDITION, WHICH HAS BEEN IDENTIFIED AS A POTENTIAL SAFETY ISSUE. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, RESULTING IN FIBER BREAKAGE AND CAP DAMAGE. THE PRODUCT LABELING WARNS THAT TISSUE CONTACT OR TISSUE PROBING MAY CAUSE FIBER DAMAGE OR BREAKAGE.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE FIBERLIFE MESSAGE WAS FLASHING CONTINUOUSLY WHILE TRYING TO VAPORIZE THE PROSTATE. THERE WAS NO REPORT OF INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANGLED DELIVERY DEVICE, GREENLIGHT XPS SURGICAL FIBER GEX AMERICAN MEDICAL SYSTEMS, INNOVATION CENTER - SILICON VALLEY 0010-2400 120A

Patients

Seq Age Sex Outcome Treatment
1 GREENLIGHT XPS SURGICAL LASER SYSTEM| ACCESSORIES