FDA Adverse Event Malfunction Summary report: N

ACCESS ESTRADIOL REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER

MDR report key: 2540121 · Received April 18, 2012

Report

Report Number
2122870-2012-01063
Event Type
Malfunction
Date Received
April 18, 2012
Date of Event
March 20, 2012
Report Date
March 20, 2012
Manufacturer
BECKMAN COULTER INC.
Product Code
CHP
PMA / PMN Number
K023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - TO CLARIFY THAT PATIENT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY USING THE NEW ANTIBODY.

Description of Event or Problem · 1

THE CUSTOMER SCREENED THE NEW REPLACEMENT ANTIBODY BEFORE REPORTING PATIENT RESULTS. THE SCREEN WAS CONDUCTED ON A 55 PATIENT COMPARATIVE STUDY BETWEEN ESTRADIOL2 LOT 117695 (NEW POLYCLONAL ANTIBODY REAGENT) AND LOT 110598. THE BIOCHEMIST REPORTED THE CORRELATION STUDIES SHOWED POSITIVE BIAS OF (B)(4) WHEN EVALUATING ACCESS ESTRADIOL REAGENT WITH NEW ANTIBODY REPLACEMENT, LOT 117695. PATIENTS RESULTS WERE REPORTED OUT OF THE LABORATORY USING THE NEW ANTIBODY. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT. THE CUSTOMER STATED THIS POSITIVE BIAS IS LARGE AND MAY IMPACT INTERPRETATION IN POST MENOPAUSAL FEMALES AND MALES. THE CUSTOMER WILL SEND PATIENT SAMPLES FOR ESTRADIOL TO ALTERNATE FACILITY TO BE TESTED UNDER ALTERNATE METHODOLOGY. CONTROLS WERE RUN BEFORE AND AFTER EVENT. THE UNIT CURRENTLY PERFORMING WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS (ACCURACY) AND REPRODUCIBILITY (PREVISION). SERVICE WAS NOT DISPATCHED FOR THIS EVENT. ROOT CAUSE IN UNKNOWN. INSTRUMENT ASSOCIATED WITH THIS EVENT IS DEVICE NAME: UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM. CLASSIFICATION CODE: ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE. CATALOG NUMBER: A30260; SERIAL NUMBER: (B)(4); SOFTWARE VERSION: 4.4; MANUFACTURE DATE: 05/07/2009.

Description of Event or Problem · 1

PATIENTS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY USING THE NEW ANTIBODY.