ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2012-01134
- Event Type
- Malfunction
- Date Received
- April 17, 2012
- Date of Event
- February 1, 2012
- Report Date
- March 20, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SYSTEM CHECK DATA HAS NOT BEEN SUPPLIED TO DATE. THERE IS NO INDICATION THAT A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT TO DATE. A DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE. THE FOLLOWING MDRS (TOTAL COUNT: 16) LISTED BELOW INCLUDES THE COMPLETE LIST OF REPORTS SUBMITTED FOR THIS EVENT THAT OCCURRED AT THIS CUSTOMER SITE: 2122870-2012-01128, 2122870-2012-01129, 2122870-2012-01130, 2122870-2012-01131, 2122870-2012-01132, 2122870-2012-01133, 2122870-2012-01134, 2122870-2012-01135, 2122870-2012-01136, 2122870-2012-01137, 2122870-2012-01138, 2122870-2012-01139, 2122870-2012-01140, 2122870-2012-01141, 2122870-2012-01142, 2122870-2012-01143.
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) IN REGARDS TO OBTAINING HIGH BETA - HUMAN CHORIONIC GONADOTROPIN (B-HCG) QC LEVEL 1 RESULTS WITH A RECOVERY OF > 2 STANDARD DEVIATION (SD) ON THE ACCESS 2 IMMUNOASSAY SYSTEM USED IN CONJUNCTION WITH TWO (2) DIFFERENT ACCESS TOTAL B-HCG REAGENTS (LOT #S 170125 AND 170067). THE ERRONEOUS QC RESULTS WERE GENERATED OVER THE COURSE OF SEVERAL DAYS RANGING FROM (B)(6) 2012. THIS REPORT DOCUMENTS THE ELEVATED QC RESULTS OBTAINED ON (B)(6) 2012. AN INVESTIGATION OF THE CUSTOMER SUPPLIED DATA INDICATES THAT THE BACKGROUND RELATIVE LIGHT UNITS (RLUS) BEGIN INCREASING AFTER A BHCG REAGENT PACK IS PUT INTO USE ON THE SYSTEM. IN COMPARING THE CUSTOMER'S DATA TO DATA FROM AN 'UNAFFECTED' REAGENT PACK SHOWS THAT THERE IS A STEADY INCREASE IN RLU AND DOSE RECOVERY AS TESTING THAT IS PERFORMED WITH AN 'AFFECTED' PACK. THIS INCREASE IN RLU MANIFESTS ITSELF AS ELEVATED L1 QC AND LOW BHCG PATIENTS' SAMPLES. IT WAS NOTED IN THE CUSTOMER DATA THAT WHEN THE CUSTOMER CHANGES TO A FRESH REAGENT PACK, THE RLUS RETURN TO THE ORIGINAL BACKGROUND LEVEL. THE DATA SHOWS THAT WHEN THE CUSTOMER CALIBRATES USING AN 'AFFECTED' PACK WITH ELEVATED RLUS, LEVEL 1 QC WILL RECOVER WITHIN RANGE AS THE CALIBRATION CURVE HAS CALCULATED THE RLU INCREASE. HOWEVER, THE SYSTEM ALLOWS MULTIPLE REAGENT PACKS TO BE LOADED FOR A SINGLE ASSAY AND THE SYSTEM CAN AUTOMATICALLY SWITCH TO A NEW 'UNAFFECTED' REAGENT PACK ONCE THE AFFECTED PACK IS EMPTY. IN THIS SITUATION, WHEN THE SYSTEM CHANGES TO A FRESH PACK WITH NORMAL BACKGROUND RLU RECOVERY, THE CALCULATED DOSE FOR SAMPLES TESTED WITH THE FRESH PACK WILL BE LOWER DUE TO THE BIAS IN THE CALIBRATION CURVE GENERATED FROM THE 'AFFECTED' PACK. THE CUSTOMER CAN POTENTIALLY OBTAIN FALSE NEGATIVE RESULTS FOR PATIENTS' SAMPLES ON THE LOWER END OF THE CALIBRATION CURVE IN SUCH A SITUATION. NO ERRONEOUS PATIENT RESULTS HAVE BEEN REPORTED TO DATE; HOWEVER, DUE TO THE ELEVATION IN ASSAY DOSE RESPONSE OVER THE USE OF AN 'AFFECTED' PACK, THERE IS POTENTIAL FOR THE CUSTOMER TO RESULT FALSE NEGATIVE RESULTS AT THE LOW END OF THE CALIBRATION CURVE. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | ACCESS® 2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |