ERBE BIPOLAR CABLE
Report
- Report Number
- 9610614-2026-00087
- Event Type
- Malfunction
- Date Received
- May 29, 2026
- Date of Event
- April 20, 2026
- Report Date
- May 29, 2026
- Manufacturer
- ERBE ELEKTROMEDIZIN GMBH
- Product Code
- GEI
- PMA / PMN Number
- K190823
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE INVOLVED BIPOLAR CABLE WAS MADE AVAILABLE AND THOROUGHLY INSPECTED. A VISUAL EXAMINATION REVEALED TEAR IN THE INSULATION IMMEDIATELY BEHIND THE KINK PROTECTION OF THE 90° ANGLED INSTRUMENT CONNECTOR. THE AREA BEHIND THE KINK PROTECTION SHOWED VISIBLE CHARRING. THE INNER CONDUCTORS AT THIS PART OF THE CABLE WERE SEVERED AND CHARRED. BASED ON THE CHARRING AND TEAR AT THIS CABLE LOCATION, THE ELECTRICAL ARC MOST LIKELY OCCURRED FROM THIS AREA. THE CABLE SHOWED TYPICAL SIGNS OF WEAR. THE CABLE IS SIX (6) YEARS OLD AND HAS UNDERGONE 76 USE/STERILIZATION CYCLES. BASED UPON THE INSPECTION OF THE CABLE, IT APPEARS THAT THE ACCESSORY WAS SUBJECTED TO FREQUENT AND INTENSE BENDING AND TENSILE STRESSES. THESE STRESSES CAUSED THE CABLE BREAK WHICH CAN PRODUCE A SHORT CIRCUIT OR VOLTAGE FLASHOVER. THIS CAN LEAD TO SPARKING OR AN ELECTRIC ARC. THE CABLE NOTES ON USE (NOU) EXPRESSLY STATES THAT THE CABLE SHOULD BE INSPECTED BEFORE USE AND NO LONGER BE USED IF IT IS DAMAGED. ALSO, IT IS RECOMMENDED IN THE NOU TO CHECK THE ELECTRICAL CONDUCTIVITY BEFORE EACH USE.
IT WAS REPORTED THAT AN INCIDENT OCCURRED WITH THE ERBE BIPOLAR CABLE DURING A HYSTERECTOMY. THE BIPOLAR CABLE WAS USED WITH A STORZ RESECTOSCOPE. NO OTHER INFORMATION WAS PROVIDED REGARDING ANY OTHER ACCESSORY OR THE ESU USED IN THE PROCEDURE. PER THE REPORT, "THE CABLE OF A BIPOLAR RESECTOSCOPE CAUSED AN ELECTRICAL ARC ON THE SURGICAL DRAPE." THIS RESULTED IN THE SURGEON'S GLOVE "MELTING." THERE WAS NO REPORTED PATIENT OR USER INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 465414 | ERBE BIPOLAR CABLE | BIPOLAR CABLE | GEI | ERBE ELEKTROMEDIZIN GMBH | 0819 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |