FDA Adverse Event Malfunction Summary report: N

FREESTYLE LIBRE 2 PLUS

MDR report key: 25333132 · Received May 29, 2026

Report

Report Number
2954323-2026-101688
Event Type
Malfunction
Date Received
May 29, 2026
Date of Event
May 4, 2026
Report Date
May 29, 2026
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
QBJ
UDI-DI
00357599871000
PMA / PMN Number
K223435
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION ACTIVITIES OR IF ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 53, 39, 89 MG/DL COMPARED TO READINGS OF 101, 168, 136 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 8 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
620767 FREESTYLE LIBRE 2 PLUS FLASH GLUCOSE MONITORING SYSTEM QBJ ABBOTT DIABETES CARE INC 78747-01 00357599871000

Patients

Seq Age Sex Outcome Treatment
1