I-STAT PT/INR CARTRIDGE
Report
- Report Number
- 2245578-2012-00278
- Event Type
- Malfunction
- Date Received
- April 3, 2012
- Date of Event
- October 17, 2011
- Report Date
- March 14, 2012
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- GJS
- PMA / PMN Number
- K020355
- Removal / Correction Number
- APOC2012-003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). AN INVESTIGATION THAT WAS COMPLETED ON (B)(4) 2012 AND A DEFICIENCY WAS IDENTIFIED. (B)(4). DETAILS WERE PROVIDED WITH THE ACTION THAT WAS CONDUCTED IN COOPERATION WITH THE US FOOD AND DRUG ADMINISTRATION.
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING PT/INR CARTRIDGES THAT YIELDED A DISCREPANT PT/INR RESULTS WHEN COMPARED TO THE LAB METHOD. CUSTOMER USED SEVERAL LOTS IN 2011. FOR 2011, THE CUSTOMER NOTED THAT 122 INR'S WERE ABOVE 4.0 OUT OF 3233 OR 3.8%. LOT#: R11243B; MFR DATE: 08/2011; EXP DATE: 02/14/2012. LOT#: R11253; MFR DATE: 09/2011; EXP DATE: 02/28/2012. LOT#: R11270; MFR DATE: 09/2011; EXP DATE: 03/14/2012. LOT#: R11272A; MFR DATE: 09/2011; EXP DATE: 03/14/2012. LOT#: R11294B; MFR DATE: 10/2011; EXP DATE: 04/14/2012. BASED ON THE INFO AVAILABLE AT THIS TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT PT/INR CARTRIDGE | PT/INR CARTRIDGE | GJS | ABBOTT POINT OF CARE INC. | NA | R11203B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |