SYNERGY MEGATRON
Report
- Report Number
- 2124215-2026-28667
- Event Type
- Malfunction
- Date Received
- May 28, 2026
- Date of Event
- October 20, 2021
- Report Date
- May 28, 2026
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NIQ
- UDI-DI
- 08714729985402
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MY
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SELVARAJ, K., YUSOF, H.M., ERNG, T. AND GHAPAR, A.K.A. (2025), CORONARY STENT LOSS RETRIEVAL USING PULMONARY BIOPSY FORCEPS VIA TRANSRADIAL APPROACH - A CASE SERIES. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 106: 1310-1314. HTTPS://DOI.ORG/10.1002/CCD.31674.
IT WAS REPORTED THAT VIA LITERATURE ARTICLE THAT STENT DISLODGEMENT OCCURRED RESULTING IN ADDITIONAL INTERVENTION. THE PATIENT PRESENTED FOR ELECTIVE PERCUTANEOUS CORONARY INTERVENTION. THE CHRONIC TOTAL OCCLUSION (CTO) TARGET LESION WAS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE CTO LESION WAS SUCCESSFULLY CROSSED AND PREDILATED WITH 3.0 MM NC AND 3.0 MM CUTTING BALLOONS. IVUS SHOWED NAPKIN-RING CALCIFICATION IN DISTAL RCA, WITH NODULAR CALCIFICATION AND DISSECTION IN MID RCA NECESSITATING TARGETED ORBITAL ATHERECTOMY OF THE DISTAL RCA FOLLOWED BY 3.5 NC BALLOON PREDILATION. DISTAL RCA WAS STENTED WITH 3.5X38 MM DES OVERLAPPING WITH A 4.0 X 28 MM DES IN PROXIMAL-MID RCA. DUE TO PROXIMAL STENT EDGE DISSECTION, A 5.0 X 20 MM SYNERGY MEGATRON EVEROLIMUS-ELUTING STENT WAS INITIALLY AIMED FOR DEPLOYMENT FROM THE OSTIAL RCA. HOWEVER, DURING DEPLOYMENT, THE PATIENT TOOK A DEEP BREATH DISLODGING THE STENT INTO THE AORTA. LOW PRESSURE STENT BALLOON INFLATION WAS ABLE TO GRIP THE STENT FOR RETRACTION INTO THE RADIAL ARTERY. LONGITUDINAL COMPRESSION OF THE LARGE 5.0 STENT WAS NOTED WITHIN RADIAL ARTERY. THE RAFR TECHNIQUE WAS SUCCESSFULLY APPLIED. DEPLOYMENT OF PULMONARY BIOPSY FORCEPS FOR COMPLETE STENT RETRIEVAL. WE THEN PROCEEDED WITH OSTIAL RCA STENTING USING THE SAME RADIAL SHEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 536086 | SYNERGY MEGATRON | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC CORPORATION | 08714729985402 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |