FDA Adverse Event Malfunction Summary report: N

SYNERGY MEGATRON

MDR report key: 25319691 · Received May 28, 2026

Report

Report Number
2124215-2026-28667
Event Type
Malfunction
Date Received
May 28, 2026
Date of Event
October 20, 2021
Report Date
May 28, 2026
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
NIQ
UDI-DI
08714729985402
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SELVARAJ, K., YUSOF, H.M., ERNG, T. AND GHAPAR, A.K.A. (2025), CORONARY STENT LOSS RETRIEVAL USING PULMONARY BIOPSY FORCEPS VIA TRANSRADIAL APPROACH - A CASE SERIES. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 106: 1310-1314. HTTPS://DOI.ORG/10.1002/CCD.31674.

Description of Event or Problem · 0

IT WAS REPORTED THAT VIA LITERATURE ARTICLE THAT STENT DISLODGEMENT OCCURRED RESULTING IN ADDITIONAL INTERVENTION. THE PATIENT PRESENTED FOR ELECTIVE PERCUTANEOUS CORONARY INTERVENTION. THE CHRONIC TOTAL OCCLUSION (CTO) TARGET LESION WAS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE CTO LESION WAS SUCCESSFULLY CROSSED AND PREDILATED WITH 3.0 MM NC AND 3.0 MM CUTTING BALLOONS. IVUS SHOWED NAPKIN-RING CALCIFICATION IN DISTAL RCA, WITH NODULAR CALCIFICATION AND DISSECTION IN MID RCA NECESSITATING TARGETED ORBITAL ATHERECTOMY OF THE DISTAL RCA FOLLOWED BY 3.5 NC BALLOON PREDILATION. DISTAL RCA WAS STENTED WITH 3.5X38 MM DES OVERLAPPING WITH A 4.0 X 28 MM DES IN PROXIMAL-MID RCA. DUE TO PROXIMAL STENT EDGE DISSECTION, A 5.0 X 20 MM SYNERGY MEGATRON EVEROLIMUS-ELUTING STENT WAS INITIALLY AIMED FOR DEPLOYMENT FROM THE OSTIAL RCA. HOWEVER, DURING DEPLOYMENT, THE PATIENT TOOK A DEEP BREATH DISLODGING THE STENT INTO THE AORTA. LOW PRESSURE STENT BALLOON INFLATION WAS ABLE TO GRIP THE STENT FOR RETRACTION INTO THE RADIAL ARTERY. LONGITUDINAL COMPRESSION OF THE LARGE 5.0 STENT WAS NOTED WITHIN RADIAL ARTERY. THE RAFR TECHNIQUE WAS SUCCESSFULLY APPLIED. DEPLOYMENT OF PULMONARY BIOPSY FORCEPS FOR COMPLETE STENT RETRIEVAL. WE THEN PROCEEDED WITH OSTIAL RCA STENTING USING THE SAME RADIAL SHEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536086 SYNERGY MEGATRON CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC CORPORATION 08714729985402

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention