FDA Adverse Event Malfunction Summary report: N

HALOGEN LIGHT SOURCE

MDR report key: 25319459 · Received May 28, 2026

Report

Report Number
3002808148-2026-18355
Event Type
Malfunction
Date Received
May 28, 2026
Report Date
May 28, 2026
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FCQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE FAN DOES NOT ROTATE DUE TO A TEMPERATURE FUSE DISCONNECTION SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE CUSTOMER RETURNED THE HALOGEN LIGHT SOURCE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THE FAN DOES NOT ROTATE DUE TO A TEMPERATURE FUSE DISCONNECTION, AND THE LAMP DOES NOT LIGHT UP. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
576363 HALOGEN LIGHT SOURCE HALOGEN LIGHT SOURCE FCQ SHIRAKAWA OLYMPUS CO., LTD. CLH-2

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown