HALOGEN LIGHT SOURCE
MDR report key: 25319459
·
Received May 28, 2026
Report
- Report Number
- 3002808148-2026-18355
- Event Type
- Malfunction
- Date Received
- May 28, 2026
- Report Date
- May 28, 2026
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FCQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE FAN DOES NOT ROTATE DUE TO A TEMPERATURE FUSE DISCONNECTION SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem · 0
THE CUSTOMER RETURNED THE HALOGEN LIGHT SOURCE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THE FAN DOES NOT ROTATE DUE TO A TEMPERATURE FUSE DISCONNECTION, AND THE LAMP DOES NOT LIGHT UP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 576363 | HALOGEN LIGHT SOURCE | HALOGEN LIGHT SOURCE | FCQ | SHIRAKAWA OLYMPUS CO., LTD. | CLH-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |