FREESTYLE LIBRE 14 DAY
Report
- Report Number
- 2954323-2026-99528
- Event Type
- Malfunction
- Date Received
- May 28, 2026
- Date of Event
- May 17, 2026
- Report Date
- May 28, 2026
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- PZE
- UDI-DI
- 05021791715374
- PMA / PMN Number
- P160030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION ACTIVITIES OR IF ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 262, 386, 391, 378 MG/DL COMPARED TO READINGS OF 110, 165, 142, 160 MG/DL RESPECTIVELY OBTAINED ON A COMPETITOR BRAND METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS 3 DAYS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83668 | FREESTYLE LIBRE 14 DAY | FLASH GLUCOSE MONITORING SYSTEM | PZE | ABBOTT DIABETES CARE INC | 71940-01 | 05021791715374 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |