FDA Adverse Event Malfunction Summary report: N

CONNECTING ROD HOFFMANN II MICRO Ø 3 X 60MM

MDR report key: 25317818 · Received May 28, 2026

Report

Report Number
0008031020-2026-00722
Event Type
Malfunction
Date Received
May 28, 2026
Date of Event
October 31, 2025
Report Date
May 28, 2026
Manufacturer
STRYKER GMBH
Product Code
KTT
UDI-DI
04546540389923
PMA / PMN Number
K033476
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT COULD BE CONFIRMED, SINCE THE PRODUCT WAS RETURNED FOR EVALUATION AND MATCHES THE ALLEGED FAILURE. THE DEVICE INSPECTION REVEALED THE FOLLOWING: VISUAL EXAMINATION OF THE RECEIVED PRODUCT SHOWS THAT BOTH ENDS HAVE FINE CRACKS, AS REPORTED. OTHERWISE, THE PART IS IN A GOOD AND NOT DAMAGED CONDITION. INSPECTION OF THE RECEIVED INFORMATION/EVIDENCE ARE DONE OR WILL BE DONE, IN THE PROCEEDINGS OF THE NC. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A DESIGN RELATED ISSUE.

Description of Event or Problem · 0

AS REPORTED: "DURING INSPECTIONS FOR PQH (B)(4) FURTHER CRACKS WERE NOTICED".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326964 CONNECTING ROD HOFFMANN II MICRO Ø 3 X 60MM APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT KTT STRYKER GMBH 26125H 04546540389923